Fabrication, testing, and clinical validation of an epoxy resin-based range shifter for pencil beam scanning proton therapy.
Kantaram Darekar, Bhushankumar J Patil, Akshay Ashok Bhavke, Umesh Bharat Gayake, Vysakh Raveendran, Rohidas Namdeo Punde, Prakash Hari Shinde, Sanjay D Dhole,
Department of Radiation Oncology, Advanced Centre for Treatment Research and Education in Cancer, Homi Bhabha National Institute, Sector 22, Kharghar, Panvel, Raigarh, Navi Mumbai Maharashtra, Navi Mumbai, MH, 410210, India. · Department of Physics, MES Abasaheb Garware College, Karve Rd, opp. Sahyadri Hospital,, Erandwane, Pune,, Maharashtra 411004, Pune, MH, 411004, India.
Key points
From the paper's conclusion- This study presents the first comprehensive clinical validation and evaluation of cured solid epoxy resin as an RS material for PBS proton therapy applications. The in-house-fabricated, cured epoxy resin-based RS demonstrated excellent agreement with TPS-MC calculations and experimental measurements, establishing its clinical feasibility.
Abstract
BACKGROUND AND OBJECTIVES: Range shifters (RS) are used to treat shallow targets in pencil beam scanning (PBS) proton therapy. This study aims to validate an in-house-fabricated epoxy resin RS through mechanical and dosimetric measurements, evaluate the Monte Carlo (MC) algorithm dose‑calculation accuracy of the treatment planning system (TPS) beam model incorporating this device, and assess its clinical feasibility for PBS proton therapy applications. METHODS: The RS was fabricated using a silicone mould and machined to the required dimensions after curing. It was further evaluated through material characterization, mechanical testing, and dosimetric validation. Material characterization included determining the elemental composition, mass density, and mean excitation energy, which were incorporated into the beam modelling. Mechanical validation comprised dimensional verification, fitment testing, and assessment of material integrity and uniformity. Dosimetric validation was performed by measuring water-equivalent thickness (WET), range parameters, spot size, absolute dose, and fluence, and comparing them with TPS-MC calculations. RESULTS: The cured epoxy resin exhibited normalized elemental composition of H:8.0%, C:65.0%, N:0.1%, and O:26.9%, with mass density 1.15 g/cm³ and mean excitation energy 72.5 eV. The fabricated RS matched the reference design within ±1 mm and demonstrated satisfactory fitment, reproducibility, material integrity and uniformity. The measured WET agreed with TPS estimated values within 0.06 g/cm², while differences in range parameters were <1 mm. Measured spot sizes and the 9-spot matrix showed excellent agreement with TPS estimates across different air-gap settings. Absolute dose measurements agreed with TPS-MC calculations within 3%, and patient-specific fluence measurements showed excellent agreement with TPS calculations under stringent gamma evaluation criteria. CONCLUSION: This study presents the first comprehensive clinical validation and evaluation of cured solid epoxy resin as an RS material for PBS proton therapy applications. The in-house-fabricated, cured epoxy resin-based RS demonstrated excellent agreement with TPS-MC calculations and experimental measurements, establishing its clinical feasibility.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of a Cable-Free and User-Friendly thermoAcoustic Range Verifier for a 2 Gy Proton Field Delivered to the Liver of an Anthropomorphic Phantom.International journal of radiation oncology, biology, physics · 2026
- Rapid commissioning of Brainlab Elements for SRS: end-to-end evaluation of Monte Carlo and pencil beam algorithm across homogeneous and heterogeneous phantom environments.Frontiers in oncology · 2026
- Proton treatment-plan dose verification with a DPM-based Monte Carlo engine.Frontiers in oncology · 2026
- Monte Carlo-based secondary dose check platform for the world's first ultra-compact upright proton therapy systemarXiv · 2026
- easyPET.3D Measurement of Positron Emitter Maps in Proton-Irradiated PMMA Phantoms.Biomedical physics & engineering express · 2026
- Megavoltage photon minibeam production through the embedded multi-leaf collimator of a clinical linac: A dosimetric characterization.Medical physics · 2026