Rapid commissioning of Brainlab Elements for SRS: end-to-end evaluation of Monte Carlo and pencil beam algorithm across homogeneous and heterogeneous phantom environments.
Bhuvaneswari Narayanan, Daniel Saenz, Nikos Papanikolaou, Neil Kirby, Doug Soltesz, Eva Galvan, Richard Tuli, Karl Rasmussen,
Department of Radiation Oncology, University of Texas at San Antonio, San Antonio, TX, United States.
Key points
From the paper's conclusion- The reference beam model provided an efficient framework for commissioning the Brainlab Elements TPS. Brainlab Elements' MC dose calculation algorithm provides reliable accuracy across clinically relevant heterogenous and small field conditions.
Abstract
PURPOSE: Brainlab released a reference beam model that accelerates Brainlab Elements treatment planning system (TPS) commissioning process for stereotactic radiosurgery (SRS) using an Elekta Versa HD linear accelerator. A series of end-to-end (E2E) phantom studies were performed to evaluate the Monte Carlo (MC) and pencil beam (PB) dose calculation algorithms using the reference beam model released by Brainlab with Elements v4.0 and v4.5. METHODS: A reduced set of water tank measurements of percent depth dose (PDD), beam profiles, multileaf collimator (MLC) gap, and output factors were taken using a PTW 0.125-cc Semiflex ionization chamber, PTW Si-diode, and PTW Microdiamond chamber for field sizes including 6 × 6, 10 × 10, 30 × 30, and 100 × 100 mm2 using 6MV and 6MV flattening filter-free (FFF) beams. MPPG 5.b testing was performed after commissioning to validate TPS performance. A set of RTsafe cranial and spine phantoms and an Exradin A16 ionization chamber was used for E2E testing. The cranial phantom was used to evaluate single target and dynamic conformal arc therapy (DCAT) cranial planning, and spine phantom filled with water, air, and air with a steel rod insert was used to evaluate spine SRS planning. MC and PB (v4.0 and v4.5) dose calculation algorithms in Brainlab Elements were used to evaluate the planning modules across all phantom environments. RESULTS: All commissioning validation tests satisfied the MPPG 5.b tolerance criteria. The commissioning process took a qualified medical physicist (QMP) and a medical physics resident 1 day to complete. MC dose calculation (2%/1 mm) demonstrated clinically acceptable agreement within 3% of measured dose for E2E tests. In-field PB calculations were within 5%. PB calculated plans showed some improvement with measured values when upgrading the beam model from v4.0 to v4.5. CONCLUSION: The reference beam model provided an efficient framework for commissioning the Brainlab Elements TPS. Brainlab Elements' MC dose calculation algorithm provides reliable accuracy across clinically relevant heterogenous and small field conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of the environ SRS immobilization mask system for HyperArc: Dosimetric attenuation, dose calculation accuracy, and clinical setup reproducibility.Journal of applied clinical medical physics · 2026
- SRS/SBRT commissioning process with a hybrid beam data model in new preconfigured CyberKnife system.Journal of applied clinical medical physics · 2026
- Evaluation of GPU-accelerated Monte Carlo beam and agility MLC modeling in the Monaco treatment planning system.Journal of applied clinical medical physics · 2026
- Beam data and model validation for Elekta SRS cone collimator in 6MV and 6MVFFF beams.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- A Compact Eight-Parameter Analytical Model of Photon-Beam Lateral Dose Profiles: Plateau, Penumbra and Out-of-Field Tail Decomposition for Elekta Linear Accelerators.Biomedical physics & engineering express · 2026
- Closing the gap: A two-site MLC matching study for Varian TrueBeam Linear AcceleratorsarXiv · 2026