Megavoltage photon minibeam production through the embedded multi-leaf collimator of a clinical linac: A dosimetric characterization.
Stefania Linsalata, Jake Harold Pensavalle, Fabiola Paiar, Francesco Romano, Antonio Claudio Traino, Fabio Di Martino
UO Fisica Sanitaria, Azienda Ospedaliero-Universitaria Pisana, Pisa, Italy. · S.I.T. - Sordina IORT Technologies S.p.A., Aprilia, Italy.
Key points
From the paper's conclusion- We are claiming that a Varian TrueBeam™ equipped with an HD-MLC represents a practical, stable, reproducible and accessible platform that might be suitable for radiobiological studies on SFRT and for the investigation of the relative underlying mechanisms.
Abstract
BACKGROUND: Although spatially-fractionated radiation therapy (SFRT) has been receiving increasing clinical interest, its complete and optimized clinical transfer requires several technological and radiobiological issues to be addressed. PURPOSE: The aim of this work is to setup and dosimetrically characterize an experimental platform generating planar megavoltage photon minibeams to exploit quantitative radiobiological SFRT experiments. METHODS: Minibeam patterns were generated on a Varian TrueBeam™ linear accelerator equipped with a high definition (HD) MLC, which consists of 60 tungsten leaf pairs, whom the 32 inner ones project a thickness of 2.5 mm at the isocenter plane (i.e., 100 cm distance from the radiation source) and the 28 outers of 5 mm. Each pattern was obtained alternating opening one leaf after closing one, two, or three of the smallest MLC leaves. Measurements were performed with a PTW microdiamond detector in edge-on orientation in a water phantom over depths from 1 to 30 cm, varying the beam energy and the source to surface distance (SSD) in different sessions. A Python-based in-house automated fitting procedure was used to obtain from the measured profiles all the minibeam parameters: peak widths (PWs), distance between consecutive peaks (CTCs), peak to valley dose ratios (PVDRs) and depth doses. RESULTS: The best performances were obtained at SSD = 70 cm and 6 MV, where PWs and PVDRs respectively range from 1.2 to 2.4 mm and from 2 to 7. For all patterns the average dose varies in depth analogously to the reference field, PVDRs show an almost steady decrease except at shallower depths, while PWs and CTCs widen linearly with the depths. Moreover, their variation over different sessions was within the estimated uncertainty of the single measurement. CONCLUSIONS: We are claiming that a Varian TrueBeam™ equipped with an HD-MLC represents a practical, stable, reproducible and accessible platform that might be suitable for radiobiological studies on SFRT and for the investigation of the relative underlying mechanisms.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Microbeam radiotherapy - clinical and dosimetric investigation of a first clinical application.Radiation oncology (London, England) · 2026
- Combining proton FLASH and spatially fractionated radiotherapy: experimental and simulation-based dosimetric characterizationPhysics in Medicine and Biology · 2026
- Magnetically enhanced peak‐to‐valley dose ratio for megavoltage photon minibeam radiotherapyMedical Physics · 2026
- A combined dose and microdosimetric modeling framework incorporating volume effects correlates with tissue sparing in proton minibeam radiotherapy.International journal of radiation oncology, biology, physics · 2026
- Feasibility of canine minibeam radiotherapy (MBRT) with a dual-robot kilovoltage x-ray system.Physics in medicine and biology · 2026
- Clinical Outcomes of Minibeam Radiotherapy for Superficial Skin and Soft Tissue Malignancies.International journal of radiation oncology, biology, physics · 2026