Native ultralipemic serum versus intravenous lipid emulsion for assessing lipemia interference in enzymatic ethanol assays: a CLSI EP07-based study.
Elmas Öğüş, Özge Sivri, Zeynep Bağlan Atalay, Naz Koçoğlu, Medine Alpdemir, Mehmet Şeneş
Department of Medical Biochemistry, University of Health Sciences, Ankara Training and Research Hospital, Ankara, Turkey. · Department of Medical Biochemistry, University of Health Sciences, Balıkesir Atatürk City Hospital, Balıkesir, Turkey.
Key points
From the paper's conclusion- This study was conducted prospectively in accordance with the Clinical and Laboratory Standards Institute (CLSI) EP07-Ed3 and C56-A guidelines.
Abstract
To evaluate the impact of lipemia on ethanol measurements and compare the interference effects of intravenous lipid emulsions (IVLE) with native ultralipemic material (NULM), focusing on clinical and forensic reliability. This study was conducted prospectively in accordance with the Clinical and Laboratory Standards Institute (CLSI) EP07-Ed3 and C56-A guidelines. Serum pools at three ethanol levels (15.53, 59.03 and 203.97 mg/dL) were prepared with a five-step triglyceride (TG) gradient (2.82-22.58 mmol/L) using NULM or IVLE. Analyses were performed on the Roche Cobas c502 analyzer (enzymatic method, 340 nm); repeated-measures analysis of variance (ANOVA) and a Clinical Laboratory Improvement Amendments (CLIA) 2024 total allowable error (TEa) of 20% were applied. NULM produced a bimodal interference pattern. At low ethanol levels, positive deviations reached +108.74% (p = 0.011); at moderate levels, negative deviations reached -70.05% (p = 0.008); at high levels, deviations reached -79.90% (p = 0.003), all exceeding the 20% CLIA threshold. IVLE was stable, with deviations between -0.47% and -11.21%, reaching significance only at the highest TG concentration (-27.42%, p = 0.037). Native lipemia causes clinically and forensically critical deviations in ethanol measurements, whereas IVLE substantially underestimates the true interference magnitude. Patient-derived NULM should be preferred in interference studies; lipemia index evaluation and ultracentrifugation are recommended for lipemic specimens.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Establishment and Application of a Quantitative Lipemia Index as a Novel Indicator for Assessing Lipemic Interference in Clinical Chemistry.Journal of clinical laboratory analysis · 2026
- Simulation Study on Preanalytical Variables in Coagulation Testing: Interference Effects and Threshold Determination of Hemolysis, Lipemia, Icterus and Insufficient Sample Volume.Clinical and applied thrombosis/hemostasis : official journal of the International Academy of Clinical and Applied Thrombosis/Hemostasis · 2026
- Assessment of Hemolysis, Icterus, and Lipemia Interference in 14 Clinical Chemistry and 5 Immunoassay Parameters.Clinical laboratory · 2026
- B-273 Study on the Application Effect of a Novel Serum Index Detection Method Based on Reagent BackgroundClinical Chemistry · 2026
- A-038 Validation of a Specialized Lipids Panel LDT for Serum Analysis Using the Roche Cobas 8000 c702 PlatformClinical Chemistry · 2026
- A-018 Diluent Evaluation Using the Diazyme Lp(a) ImmunoassayClinical Chemistry · 2026