A-018 Diluent Evaluation Using the Diazyme Lp(a) Immunoassay
Alexandria Jackson, Jonathan Genzen, Joseph Rudolf, Sara Wyness
Key points
From the paper's conclusion- The results of the manual dilutions and the onboard auto-dilutions were compared.
Abstract
Background The Diazyme lipoprotein(a) [Lp(a)] immunoassay is intended for measuring Lp(a) concentrations in serum or plasma. Evaluation of specimens with concentrations above the analytical measuring range (AMR) may be warranted for patients. Validation of dilution protocols (including manual dilations and/or auto-dilutions) is important to demonstrate linearity in the extended measuring interval (EMI). The present study was conducted to validate dilution ratios and diluents using this assay. Methods Residual patient samples with Lp(a) concentrations near the assay upper limit of quantitation (ULOQ; 100 mg/dL) were retrieved and de-identified using an IRB-approved protocol and manually diluted using different candidate diluents and dilution ratios. Undiluted and diluted samples were analyzed using the Diazyme Lp(a) assay installed on AU5800 chemistry analyzers (Beckman Coulter). Acceptable limits were defined as ±20% of expected sample dilution recovery. Percent sample dilution recovery was calculated as [(average(diluted sample x dilution factor))/average(undiluted sample)] x 100. Results were analyzed following guidelines outlined in CLSI EP34-Ed1:2018 using Microsoft Excel. The dilution notation s:t was used, with s = relative volume of neat specimen; and t = total volume after the diluent has been added (e.g., a 1:2 dilution uses one unit volume of specimen with one unit volume of diluent to produce two unit volumes of diluted sample). Seven diluents were evaluated. Results Of the diluents tested, only Diluent 2 (Beckman Coulter; sold for IMMAGE immunochemistry systems) was within ±20% of target recovery. All other diluents resulted in less than 80% sample dilution recovery. To determine an optimal dilution ratio for Diluent 2, samples near the ULOQ were manually diluted at a 1:10 ratio. Dilution recoveries ranged between 89.1% – 100.8%. The mean of sample percent recoveries for the 1:5 dilution ratio and 1:10 dilution ratio was 86.8% and 96.2%, respectively. Auto-dilution was then evaluated. Diluent 2 was loaded on the AU5800 and five samples with Lp(a) concentrations spanning the EMI were manually diluted and tested using Diluent 2 at 1:10 dilution ratio, as well as analyzed using auto-dilution programmed at a 1:10 ratio. The results of the manual dilutions and the onboard auto-dilutions were compared. Sample dilution recoveries using 1:10 auto-dilution ranged from 97.9% - 100.9%. Conclusions Dilution studies identified both a diluent and dilution ratio that met pre-defined acceptability criteria. The dilution ratio was also compatible with instrument auto-dilution functionality to improve laboratory workflow and efficiency. Laboratories should validate dilution ratios and diluents, as appropriate, to exclude non-linearity across the EMI.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Harmonization of Lipoprotein(a) Immunoassays with the IFCC- Endorsed Reference Measurement Procedure Validated in Patient Samples and Multiple Routine testing as A Second Step Towards Apolipoprotein Standardization.Journal of atherosclerosis and thrombosis · 2026
- Analytical validation of a direct lipoprotein(a)-cholesterol assayJournal of Lipid Research · 2026
- B-300 Validation of the Lp(a) Ultra Assay on Beckman Coulter AU680: Overcoming Isoform Heterogeneity Challenges in Lipoprotein(a) MeasurementClinical Chemistry · 2026
- A-017 Identifying and Resolving Outliers in Lipoprotein(a) Immunoassay TestingClinical Chemistry · 2026
- Lipoprotein(a) analytical variation and cardiovascular thresholds.Clinical chemistry and laboratory medicine · 2026
- Establishment and Application of a Quantitative Lipemia Index as a Novel Indicator for Assessing Lipemic Interference in Clinical Chemistry.Journal of clinical laboratory analysis · 2026