A turbidimetric alternative for analysis of total light chains in urine.
Joakim Pejer, Annika Petersson, Kim Ekblom, Ulrik Hägglund, Magnus Förnvik Jonsson
Department of Clinical Chemistry and Transfusion Medicine, Centrallasarettet Vaxjo, Växjö, Sweden. · Department of Clinical Chemistry and Pharmacology, Skåne University Hospital, Lund, Sweden.
Key points
From the paper's conclusion- We evaluated Siemens N Antiserum (N AS) for total kappa chains and total lambda chains on a cobas 502 turbidimeter (Roche Diagnostics).
Abstract
Commercial assays quantifying total immunoglobulin light chains in urine are generally intended for nephelometry but are used on turbidimetric analyzers. We evaluated Siemens N Antiserum (N AS) for total kappa chains and total lambda chains on a cobas 502 turbidimeter (Roche Diagnostics). Imprecision was calculated at several levels based on repetitive measurements of controls, patient samples, and calibrator solutions. To approximate functional sensitivity, a precision profile was generated. Bias was estimated using 150 patient urine samples, both on the cobas and a BNII nephelometer (Siemens Healthineers). The comparisons were performed using Passing-Bablok regressions and Bland-Altman plots. Our results showed a CV <10% at concentrations >10 mg/L for both the kappa and lambda assays. A CV of approximately 20% (functional sensitivity) was seen at approximately 5 mg/L for both methods. The cobas and BNII analyzers showed good agreement, although kappa measurements on the cobas gave slightly higher values at low concentrations. For the lambda assay, regression analysis showed no clinically relevant difference. In summary, we found no major difference between Cobas c502 and BNII. Analysis with N AS Kappa and Lambda on Cobas c502 gives clinically acceptable results at concentrations ≥8 mg/L and ≥5 mg/L, respectively.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multicenter evaluation of analytical performance and onboard stability of Mindray 10 concentrated biochemical reagentsJournal of Laboratory Medicine · 2026
- B-112 Method Comparison of the Binding Site and Diazyme Serum Free Light Chain Assays on the Roche cobas c502Clinical Chemistry · 2026
- Urinary creatinine estimation as an important analytical limitation of semi-quantitative albumin-to-creatinine ratio strip testing: a diagnostic accuracy study.Frontiers in medical technology · 2026
- B-001 Limitations of Urine Dipstick Testing and the Clinical Value of Creatinine-Normalized Quantitative Urinalysis in Urologic and Renal Disease DetectionClinical Chemistry · 2026
- Consistency and correlation of a domestic Kehua serum free light chain assay with two imported analytical systems.Frontiers in oncology · 2026
- A-372 Assessment of Urinary Albumin to Creatinine Ratio Using the Capitainer Dip70® Dried Urine Microsampling Device on Automated Clinical AnalyzersClinical Chemistry · 2026