Effect of cyclic loading on screw loosening in fixed partial dentures supported by the OT bridge versus multi-unit angled abutments: An in vitro study.

Ahmed M Abdelhamid, Mohamed Nawar, Lucette Segaan, Lubna El-Didi, Kenda I Hanno

Professor, Department of Prosthodontics, Faculty of Dentistry, University of Alexandria, Alexandria, Egypt. · Researcher, Faculty of Dentistry, University of Alexandria, Alexandria, Egypt.

Preclinical

Key points

From the paper's conclusion
  • Within this study's limitations, screw stability was comparable between groups at the premolar site. However, the OT Bridge system exhibited significantly higher screw loosening and RTL difference ratios at the second molar site compared with MUA (P<.05). While cyclic loading did not significantly trigger loosening within groups, the site-specific differences suggested that the OT Bridge design may be more sensitive to the higher mechanical demands of the molar region. Clinical monitoring in high-stress areas is recommended.

Abstract

STATEMENT OF PROBLEM: The use of angled abutments may be the method of choice when anatomic limitations preclude the axial placement of an implant. However, the most common complication in tilted implants is abutment screw loosening. PURPOSE: The purpose of this in vitro study was to measure and compare the amount of screw loosening before and after dynamic cyclic loading in the OT bridge abutment and angled multi-unit abutment in implant-supported 3-unit fixed partial dentures (FPDs). MATERIAL AND METHODS: One partially edentulous mandibular model (Class II Kennedy Classification) was used in this study. One implant was placed in the second premolar region, and 1 distally tilted implant was placed in the second molar region. A screw-in abutment was connected to the straight abutment in both groups. The OT bridge was connected to distally tilted implants in the OT bridge group, and a multi-unit angled abutment was used for the tilted implant in the multi-unit group. Sixteen 3-unit fixed partial dentures were designed and milled and allocated to 2 groups according to the type of attachment. The prosthetic screw removal torque was measured before and after dynamic cyclic loading of 100 000 cycles. The Mann-Whitney U test was used to compare groups, and the Wilcoxon Sign Rank test was used to compare values before and after loading (α=.05). RESULTS: No significant difference was found in the percentage of screw loosening at the second premolar in both groups (P=.109 and P=.136, respectively). The percentage of screw loosening in the OT bridge group at the second molar was 17.35%, which was higher than in the MUA group (16.50%) before cyclic loading, and 26.50% and 18.30% respectively after cyclic loading (P<.05). However, cyclic loading had no significant effect on screw loosening in either group (P>.05). The retentive torque loss (RTL) difference ratio was significantly higher in the OT bridge group than in the MUA group and in the second molar than in the premolar in both groups. CONCLUSIONS: Within this study's limitations, screw stability was comparable between groups at the premolar site. However, the OT Bridge system exhibited significantly higher screw loosening and RTL difference ratios at the second molar site compared with MUA (P<.05). While cyclic loading did not significantly trigger loosening within groups, the site-specific differences suggested that the OT Bridge design may be more sensitive to the higher mechanical demands of the molar region. Clinical monitoring in high-stress areas is recommended.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers