The impact of abutment manufacturing techniques on reverse torque after cyclic loading and implant-abutment interface fitting: an in vitro study.
Alaa Mostafa AboRas, Ahmed Ismail Taha, Nourhan Ahmed Ragheb
Prosthodontic Department, Faculty of Dentistry, Kafr Al Sheikh University, Mubark Road, Kafr Al Sheikh, 33511, Egypt. aborasalaa874@gmail.com. · Prosthodontic Department, Faculty of Dentistry, Kafr Al Sheikh University, Mubark Road, Kafr Al Sheikh, 33511, Egypt.
Key points
From the paper's conclusion- Within the limitations of this study, prefabricated abutments demonstrated the highest reverse torque values after cyclic loading. 3D-printed abutments showed significantly greater resistance to torque loss compared to CAD-CAM-milled abutments, approaching the performance of prefabricated abutments. The implant-abutment fit was comparable across groups, suggesting that connection design may be more influential than manufacturing method for macroscopic fit. 3D-printed titanium abutments demonstrated significantly greater mechanical stability than CAD-CAM-milled abutments, approaching the performance of the gold-standard prefabricated abutments. This suggests that additive manufacturing may offer a clinically viable alternative to milling when both customization and mechanical integrity are required. However, clinical studies are needed to confirm these in vitro findings.
Abstract
BACKGROUND: Screw loosening and implant-abutment interface (IAI) are critical for the long-term success of implant-supported restorations. This study aimed to evaluate the effect of abutment manufacturing technique on reverse torque values and micro-gap measurements at the implant-abutment interface. METHODS: Thirty dental implant-abutment-crown complexes (Implance, Verdent, Turkey) were randomly allocated into three groups (n = 10 each) based on the abutment manufacturing technique: Group I-prefabricated titanium abutments (manufacturer-supplied, Implance, Verdent, Turkey; titanium grade 5); Group II-CAD-CAM-milled titanium abutments; Group III-3D-printed titanium abutments (Selective Laser Melting, SLM). All abutments featured an internal hexagonal connection with platform switching. All crown abutments were tightened to 25 Ncm and subjected to cyclic loading (100 N, 1.6 Hz, 250,000 cycles). Removal torque was measured before and after loading. Micro-gaps at the implant-abutment and abutment-screw interfaces were evaluated using scanning electron microscopy (SEM). Statistical analysis was performed using One-Way ANOVA with Tukey's Post Hoc test for reverse torque and the Kruskal-Wallis test for gap measurements (α = 0.05). RESULTS: After cyclic loading, the 3D-printed (SLM) abutments exhibited significantly higher reverse torque values (11.08 ± 0.11 Ncm) compared to CAD-CAM-milled abutments (6.40 ± 0.79 Ncm) (P < 0.001). Prefabricated abutments demonstrated the highest post-loading torque values (13.31 ± 0.31 Ncm). All intragroup comparisons showed statistically significant reductions from baseline (P < 0.0001). No significant difference was found in implant-abutment interface gaps among groups (P = 0.105). CONCLUSION: Within the limitations of this study, prefabricated abutments demonstrated the highest reverse torque values after cyclic loading. 3D-printed abutments showed significantly greater resistance to torque loss compared to CAD-CAM-milled abutments, approaching the performance of prefabricated abutments. The implant-abutment fit was comparable across groups, suggesting that connection design may be more influential than manufacturing method for macroscopic fit. 3D-printed titanium abutments demonstrated significantly greater mechanical stability than CAD-CAM-milled abutments, approaching the performance of the gold-standard prefabricated abutments. This suggests that additive manufacturing may offer a clinically viable alternative to milling when both customization and mechanical integrity are required. However, clinical studies are needed to confirm these in vitro findings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Influence of different abutment materials on torque maintenance and screw loosening in implant-supported prostheses.Bioinformation · 2026
- Effect of Tightening Protocol on Reverse Torque and Plastic Deformation of Dynamic and Conventional Titanium Abutment Screws.Bioengineering (Basel, Switzerland) · 2026
- Effect of cyclic loading on screw loosening in fixed partial dentures supported by the OT bridge versus multi-unit angled abutments: An in vitro study.The Journal of prosthetic dentistry · 2026
- A comparative analysis of efficacy of torque limiting devices provided by different dental implant manufacturersBMC Oral Health · 2026
- Comparative analysis of stability and removal torque of custom CAD-CAM versus stock abutments before and after cyclic loading: A systematic review with meta-analysis and GRADE analysis.The Journal of prosthetic dentistry · 2026
- Internal fit of cobalt-chromium and zirconia implant abutments fabricated using different techniques: a micro-computed tomography study.International journal of implant dentistry · 2026