Stability analysis of stock alloplastic temporomandibular joint prostheses based on clinical cases of screw loosening.
Ruoyi Ding, Jiale Hong, Wendong Xue, Chuan Lu, Jieyun Zhao, Ning Wang, Dongmei He
Department of Oral Surgery, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, College of Stomatology, Shanghai Jiao Tong University, National Center for Stomatology, National Clinical Research Center for Oral Diseases, Shanghai Key Laboratory of Stomatology, Shanghai Research Institute of Stomatology, Shanghai, 200011, China. · Shanghai Key Laboratory of Orthopaedic Implants, Department of Orthopaedic Surgery, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200011, China.
Key points
From the paper's conclusion- These findings suggest that increasing component malposition progressively compromises fixation safety.
Abstract
Screw loosening is one of the common complications that affect stability after stock alloplastic temporomandibular joint (TMJ) replacement. Based on clinically observed cases of screw loosening, this study aimed to evaluate how prosthesis installation patterns influence screw loading and to provide recommendations for installation safety. A three-dimensional finite element model of unilateral total TMJ replacement was constructed to analyze Biomet standard TMJ prosthesis stability under the following situations: 1) fossa component installed at different angles (0°-20°) relative to the bone surface; 2) mandibular component installed at different angles (0°-15°) from the posterior border of the ramus. Under maximum occlusal loading and mouth-opening muscle loading, the characteristics of stress and strain, force on the fixation screws and displacement were analysed. The ratio of the maximum screw pull-out strength from the bone tested in vitro to the force on the screws was calculated to predict the safe installation angle. As the angle between the fossa component and the bone surface increased, the stress, strain and force on the prosthesis and fixation screws gradually enlarged especially the fossa screws increased by 3.05 to 9.36 times. As the angle between the mandibular prosthesis and the posterior border of the ramus increased, the stress, strain and force on the prosthesis and screws enlarged especially the fossa screws increased by 2.47 times to 4.54 times. A reduction in fixation safety was already observed at an installation angle of 5°, with individual fossa screws showing FoS values below 4. As the installation angle further increased, screw loading increased and the number of screws with reduced safety margins increased. These findings suggest that increasing component malposition progressively compromises fixation safety. The fossa component should therefore be adapted as closely as possible to the bone surface, and the mandibular component should be positioned as parallel as possible to the posterior border of the ramus.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biomechanical Evaluation of Patient-Specific Titanium Fixation Systems for Segmental Le Fort I Osteotomy: A Finite Element Analysis.Journal of functional biomaterials · 2026
- Effect of cyclic loading on screw loosening in fixed partial dentures supported by the OT bridge versus multi-unit angled abutments: An in vitro study.The Journal of prosthetic dentistry · 2026
- 20-year experience of biomet stock alloplastic temporomandibular joint replacement in Chinese patients after technical modifications: a single-center retrospective study.Clinical oral investigations · 2026
- Biomechanical and fatigue analysis of a porous titanium total temporomandibular joint prosthesis with different mandibular component lengths and designs.Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery · 2026
- Towards minimally invasive prosthetic surgery of the temporomandibular joint: design and integrated CAD-FEM preclinical approach on the mini-pig.Computer methods in biomechanics and biomedical engineering · 2026
- Optimizing Miniscrew Stability: A Finite Element Study of Titanium Screw Insertion AnglesBiomimetics · 2025