Physics and imaging in radiation oncology

Feasibility of ultra-low-field MRI in an ion beam treatment room for online MR-guided particle therapy.

Julia Pfitzer, Hermann Scharfetter, Panagiotis Antypas, Dietmar Georg, Hermann Fuchs

Institute of Biomedical Imaging, Graz University of Technology, Stremayrgasse 16/III, Graz, 8010, Austria. · Department of Radiation Oncology, Medical University of Vienna, Spitalgasse 23, Vienna, 1090, Austria.

Key points

From the paper's conclusion
  • ULF-MRI operation in an ion-beam therapy environment without a dedicated shielded room is feasible, without compromising imaging, beam delivery, or dosimetry. This approach supports future MR-guided particle therapy research with reduced infrastructural requirements.

Abstract

BACKGROUND AND PURPOSE: Online MR-guided particle therapy may improve beam delivery precision and treatment efficacy. Ultra-low-field (ULF) MRI may enable image guidance with minimal modification of existing facilities. The feasibility of operating a ULF magnetic resonance (MR) scanner for online MR-guided particle therapy in an unmodified ion-beam therapy environment was investigated. MATERIAL AND METHODS: A portable 49 mT ULF-MRI system (OSI2 ONE MR scanner) was placed at the isocenter of a research room at an ion-beam therapy facility without radio-frequency shielding. Image quality was assessed using a grid phantom under static and dynamic irradiation conditions for protons and carbon ions. Electromagnetic interference (EMI) was characterized and mitigated via grounding and local shielding. Beam deflections were quantified using a scintillation detector and compared to numerical simulations. The impact of ongoing proton and carbon ion irradiation on the imaging process was assessed. Dosimetric measurements were performed using a Semiflex ionization chamber in a PMMA phantom with and without simultaneous MR imaging. RESULTS: Stable image quality was achieved across proton and carbon ion irradiation with no relevant irradiation-induced artifacts. Initially substantial EMI was reduced to near-baseline levels by internal shielding. Proton beam deflections ranged from 0.9-1.7 mm at isocenter, consistent with simulations. Dosimetric deviations were below 0.05% for 2D fields and 0.02% for spread-out Bragg peaks (SOBP). CONCLUSION: ULF-MRI operation in an ion-beam therapy environment without a dedicated shielded room is feasible, without compromising imaging, beam delivery, or dosimetry. This approach supports future MR-guided particle therapy research with reduced infrastructural requirements.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers