Feasibility of ultra-low-field MRI in an ion beam treatment room for online MR-guided particle therapy.
Julia Pfitzer, Hermann Scharfetter, Panagiotis Antypas, Dietmar Georg, Hermann Fuchs
Institute of Biomedical Imaging, Graz University of Technology, Stremayrgasse 16/III, Graz, 8010, Austria. · Department of Radiation Oncology, Medical University of Vienna, Spitalgasse 23, Vienna, 1090, Austria.
Key points
From the paper's conclusion- ULF-MRI operation in an ion-beam therapy environment without a dedicated shielded room is feasible, without compromising imaging, beam delivery, or dosimetry. This approach supports future MR-guided particle therapy research with reduced infrastructural requirements.
Abstract
BACKGROUND AND PURPOSE: Online MR-guided particle therapy may improve beam delivery precision and treatment efficacy. Ultra-low-field (ULF) MRI may enable image guidance with minimal modification of existing facilities. The feasibility of operating a ULF magnetic resonance (MR) scanner for online MR-guided particle therapy in an unmodified ion-beam therapy environment was investigated. MATERIAL AND METHODS: A portable 49 mT ULF-MRI system (OSI2 ONE MR scanner) was placed at the isocenter of a research room at an ion-beam therapy facility without radio-frequency shielding. Image quality was assessed using a grid phantom under static and dynamic irradiation conditions for protons and carbon ions. Electromagnetic interference (EMI) was characterized and mitigated via grounding and local shielding. Beam deflections were quantified using a scintillation detector and compared to numerical simulations. The impact of ongoing proton and carbon ion irradiation on the imaging process was assessed. Dosimetric measurements were performed using a Semiflex ionization chamber in a PMMA phantom with and without simultaneous MR imaging. RESULTS: Stable image quality was achieved across proton and carbon ion irradiation with no relevant irradiation-induced artifacts. Initially substantial EMI was reduced to near-baseline levels by internal shielding. Proton beam deflections ranged from 0.9-1.7 mm at isocenter, consistent with simulations. Dosimetric deviations were below 0.05% for 2D fields and 0.02% for spread-out Bragg peaks (SOBP). CONCLUSION: ULF-MRI operation in an ion-beam therapy environment without a dedicated shielded room is feasible, without compromising imaging, beam delivery, or dosimetry. This approach supports future MR-guided particle therapy research with reduced infrastructural requirements.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The impact of electron streaming in head and neck cancer patients treated with ultra-hypofractionated MR-guided radiotherapy.Physics and imaging in radiation oncology · 2026
- MRI-only workflow for preclinical ion beam therapy: Deep-learning-based synthetic CT generation from ultra-high-field small animal MRI.Zeitschrift fur medizinische Physik · 2026
- Dosimetric validation of proton beam deflection and field shape deformation in a measured transverse magnetic field of an in-beam MR scanner using a novel dose calculation functionality in a research version of a commercial treatment planning system.Physics in medicine and biology · 2026
- EP09.29 Image-guided intraoperative radiotherapy for resectable brain metastases: Prospective evaluation of dosimetric accuracy and clinical outcomesNeuro-Oncology · 2026
- Implementation of a 0.55T MR simulator for intracranial stereotactic radiosurgeryJournal of Applied Clinical Medical Physics · 2026
- Magnetic-Field-Induced Secondary-Electron Effects in MR-Guided Radiotherapy: From Dosimetric Mechanisms to Workflow-Based Clinical Risk Management.Technology in cancer research & treatment · 2026