MRI-only workflow for preclinical ion beam therapy: Deep-learning-based synthetic CT generation from ultra-high-field small animal MRI.

Isselmou Abdarahmane, Silvia Stocchiero, Lorenz Wolf, Gerd Heilemann, Barbara Knäusl, Lukas Zimmermann, Markus Zeilinger, Dietmar Georg,

Department of Radiation Oncology, Medical University of Vienna, Währinger Gürtel 18-20, A-1090 Vienna, Austria; Competence Center for Preclinical Imaging and Biomedical Engineering, Faculty of Health, University of Applied Sciences, Johannes-Gutenberg-Straße 3, A-2700 Wiener Neustadt, Austria; MedAustron Ion Therapy Center, Marie Curie-Straße 5, A-2700 Wiener Neustadt, Austria. · Department of Radiation Oncology, Medical University of Vienna, Währinger Gürtel 18-20, A-1090 Vienna, Austria; MedAustron Ion Therapy Center, Marie Curie-Straße 5, A-2700 Wiener Neustadt, Austria.

Key points

From the paper's conclusion
  • This study demonstrated the feasibility of MRI-only treatment planning for preclinical brain irradiation using a U-Net-based sCT generator, reducing the need for dedicated µCT acquisitions. However, limited training data due to ethical constraints challenges cross-cohort generalization, while a rigidly aligned sCTavg provided competitive performance but requires an additional registration step.

Abstract

PURPOSE: To develop and validate a deep-learning-based magnetic resonance imaging (MRI)-only workflow for preclinical mouse brain irradiations with ion beams and to compare with a rigidly aligned population-averaged CTs (sCTsavg). METHODS: A 3D U-Net was trained on 69 paired ultra-high-field (15.2T) T1-weighted MRIs and micro-CTs (µCTs) from 17 BALB/c mice. Two BALB/c and two C57BL/6 mice were used for independent testing. Two training losses were compared and selected based on cross-validation. Quantitative performance was assessed using mean absolute error (MAE), structural similarity index (SSIM), and bone Dice similarity coefficient (DSC). Treatment plans with X-rays, protons, and carbon ions were analyzed using dose-volume parameters and 3D gamma analysis of the physical dose distribution. RESULTS: For image generation, the intra-strain test set reached an MAE of (138±10)HU and a bone DSC of 0.75-0.80, slightly below the sCTavg baseline (0.79-0.80). On the inter-strain test set (C57BL/6), the results were inferior (DSC: 0.59--0.64), while the sCTavg retained a DSC of 0.70-0.77. The dose distributions were comparable between plans calculated on the U-Net-based synthetic CT (sCTUNet) and on the µCT. The largest deviation occurred in the near-maximum bone dose (D2%) with 0.16Gy for both, protons and for carbon ions. CONCLUSIONS: This study demonstrated the feasibility of MRI-only treatment planning for preclinical brain irradiation using a U-Net-based sCT generator, reducing the need for dedicated µCT acquisitions. However, limited training data due to ethical constraints challenges cross-cohort generalization, while a rigidly aligned sCTavg provided competitive performance but requires an additional registration step.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers