Dosimetric validation of proton beam deflection and field shape deformation in a measured transverse magnetic field of an in-beam MR scanner using a novel dose calculation functionality in a research version of a commercial treatment planning system.
Marisa Cobanaj, Franciska Lebbink, Sergej Schneider, Louisa Schaefer, Krishna Jeanne Godino Padre, Stefan Menkel, Jörg Pawelke, Jakob Ödén,
Faculty of Medicine and University Hospital Carl Gustav Carus, TUD Dresden University of Technology, Helmholtz-Zentrum Dresden-Rossendorf, OncoRay - National Center for Radiation Research in Oncology, Fetscherstr. 74,, PF 41, Dresden, 01307, Germany. · Faculty of Medicine and University Hospital Carl Gustav Carus, TUD Dresden University of Technology, Helmholtz-Zentrum Dresden-Rossendorf, OncoRay - National Center for Radiation Research in Oncology, Fetscherstr. 74, PF 41, Dresden, 01307, Germany.
Key points
From the paper's conclusion- This study experimentally validates Monte Carlo-based proton dose calculation in the presence of the MF in a research version of the commercial TPS RayStation for MRiPT, using a measured 3D magnetic vector field map of a 0.32 T in-beam MR scanner combined with a horizontal proton pencil beam scanning beam line.
Approach.
An MR-compatible mapping framework was used to acquire a high-resolut
Abstract
Integrating in-beam magnetic resonance imaging (MRI) with proton therapy (PT) introduces magnetic fields (MFs) that affect proton beam transport and therefore require accurate incorporation into the treatment planning system (TPS) for magnetic resonance (MR)-integrated PT (MRiPT). This study experimentally validates Monte Carlo-based proton dose calculation in the presence of the MF in a research version of the commercial TPS RayStation for MRiPT, using a measured 3D magnetic vector field map of a 0.32 T in-beam MR scanner combined with a horizontal proton pencil beam scanning beam line. Approach. An MR-compatible mapping framework was used to acquire a high-resolution 3D magnetic vector field map, subsequently implemented into the TPS following coordinate transformations. Proton spot patterns and square radiation fields at 100, 150, and 220 MeV were delivered with and without the MF present. Dose distributions in air were measured using a 2D ionisation chamber array at multiple positions and compared with TPS dose calculations by analysing lateral beam shifts and field shape deformations with local γ-analysis. Clinically relevant patient treatment fields were evaluated at different water-equivalent depths (WEDs) in a phantom. Main results. Implementation of the measured magnetic vector field map preserved field vector magnitude, orientation and spatial alignment within the TPS. Dose measurements showed energy- and position-dependent MF-induced effects, which were accurately reproduced by the TPS. Agreement in lateral beam shifts was within 1-3 mm, and γ-pass rates exceeded 97% for a 3%/3 mm local criterion (10% dose threshold). Similar agreement was observed for patient treatment fields throughout all WEDs. Significance. This work validates the TPS capability to accurately model MF-induced proton beam transport in air as well as in a water-equivalent phantom. The present workflow establishes a foundation for proton treatment planning in the presence of MFs and thereby contributes to the clinical implementation of MRiPT.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Spot-wise compensation of proton beam deflection and radiation field deformation in the transverse magnetic field of an in-beam MR scanner using a research version of a commercial treatment planning system.Physics in medicine and biology · 2026
- Development and Experimental Validation of an Energy Layer-wise Beam Deflection Compensation Method for MR-integrated Proton Therapy with a Transverse Magnetic Field.International journal of radiation oncology, biology, physics · 2026
- Characterizing magnetic field coupling in MRI-guided radiotherapy by solving a Dirichlet problem with sparsely sampled boundary conditions.Physics in medicine and biology · 2026
- Magnetic-Field-Induced Secondary-Electron Effects in MR-Guided Radiotherapy: From Dosimetric Mechanisms to Workflow-Based Clinical Risk Management.Technology in cancer research & treatment · 2026
- Impact of magnetic field orientation and strength on dose distributions in 6 MV photon beams: A Monte Carlo study.Radiation and environmental biophysics · 2026
- Novel method of measuring magnetically induced ForcesCurrent Directions in Biomedical Engineering · 2026