Implementation of a 0.55T MR simulator for intracranial stereotactic radiosurgery
Joshua Kim, Robert Rusnac, Amanda Dicarlo, Anthony Doemer, Ralph Ruiz, M. Salim Siddiqui, M. Shah, Eric A. Mellon,
Key points
From the paper's conclusion- We describe the implementation of a 0.55T MR simulator and characterize lesion conspicuity across 102 lesions in 13 patients with multiple brain metastases, drawn from 211 consecutive SRS patients treated to 511 lesions over 12 months.
Abstract
Background Low‐field MRI has several advantages for radiation therapy simulation. The reduced field strength lowers susceptibility artifacts near metallic implants, decreases intrinsic geometric distortion, and improves safety through lower SAR, reduced projectile risk, and fewer device interactions. MRI is the imaging foundation for SRS target delineation, and initial low‐field MRI feasibility for brain radiotherapy planning has been demonstrated in small cohorts. However, SRS clinical implementation has not been reported. We describe the implementation of a 0.55T MR simulator and characterize lesion conspicuity across 102 lesions in 13 patients with multiple brain metastases, drawn from 211 consecutive SRS patients treated to 511 lesions over 12 months. Purpose To describe the clinical implementation of a 0.55T MR simulator (MAGNETOM Free.Max RT Edition, Siemens Healthineers) as a dedicated radiation therapy simulation system, following a systematic framework of acceptance testing per AAPM TG‐284, longitudinal quality assurance, and SRS MRI protocol optimization. Methods Acceptance testing and commissioning followed AAPM TG‐284 recommendations. Testing included the use of vendor‐provided homogeneous phantoms, the ACR large phantom, and a commercially available large‐field geometric distortion phantom. Longitudinal quality assurance was performed over a 12‐month period using the ACR large phantom. Measured parameters included geometric accuracy, percent image uniformity (PIU), percent signal ghosting, low contrast detectability (LCD), and SNR. Contrast to noise ratio (CNR) was calculated for the lesions of a subset of SRS patients having 4 or greater metastatic lesions to evaluate the conspicuity of small SRS lesions in this challenging environment. Results All acceptance testing met TG‐284 recommendations. Central frequency and transmitter gain were stable with changes of less than 0.01% and 0.3%, respectively, over the first month of testing. Geometric distortion after vendor 3D corrections measured less than 1 mm within 10 cm and less than 2 mm within 20 cm of isocenter. Monthly phantom testing over 12 months confirmed geometric stability within established tolerances. A brain imaging protocol achieved 1 mm slice thickness on 3D T1‐weighted acquisitions consistent with ISRS guidelines, with total scan time under 35 minutes. To date, 211 intracranial SRS patients have been scanned and treated to a total of 511 lesions with a median volume of 0.58cc [0.004 – 304.48 cc]. The median CNR for the 102 lesions in the set of 13 patients with 4 or greater metastatic lesions was calculated as 6.81 [1.7–16.9] with only two lesions having a CNR < 2. Conclusions These findings support the feasibility of 0.55T MRI as a dedicated RT simulator capable of meeting the imaging demands of intracranial SRS. Formal comparative studies against higher field systems are needed before low‐field imaging can be considered a standalone replacement for SRS planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- ACPSEM guideline: commissioning and quality assurance of Magnetic Resonance Imaging Linear Accelerators.Physical and engineering sciences in medicine · 2026
- Characterizing magnetic field coupling in MRI-guided radiotherapy by solving a Dirichlet problem with sparsely sampled boundary conditions.Physics in medicine and biology · 2026
- Characterizing electronic noise interference from the external laser positioning system on a radiation therapy MRI simulation scanner.Journal of applied clinical medical physics · 2026
- Magnetic Resonance Imaging Simulation of the Skull Base, Oral Cavity, Pharynx, and Larynx: Advanced Imaging Protocol for a Radiation Oncology Center.medRxiv : the preprint server for health sciences · 2026
- Feasibility of PI-RADS Compliant Prostate MRI at 0.55TmedRxiv · 2026
- From Repeatability to Reproducibility in Quantitative Low-Field MRI: A Metrology-Informed Framework for Phantom Evidence.Journal of imaging · 2026