EP09.29 Image-guided intraoperative radiotherapy for resectable brain metastases: Prospective evaluation of dosimetric accuracy and clinical outcomes

C S Dejonckheere, G R Sarria, M Grimmer, M Hamed, M Schneider, J Scorzin, V Borger, L L Friker,

Observational · n=19

Key points

From the paper's conclusion
  • We conducted a prospective study integrating intraoperative image guidance with Monte Carlo-based dosimetric reconstruction to evaluate workflow feasibility and geometric accuracy of low-energy X-ray IORT.

Abstract

Background Intraoperative radiotherapy (IORT) for resectable brain metastases (BMs) enables immediate irradiation of the surgical cavity and rapid initiation of systemic therapy. However, prospective verification of actual dose delivery is lacking. We conducted a prospective study integrating intraoperative image guidance with Monte Carlo-based dosimetric reconstruction to evaluate workflow feasibility and geometric accuracy of low-energy X-ray IORT. Material and Methods All consecutive patients undergoing resection of a supratentorial BM amenable to IORT were screened for inclusion. After IORT applicator placement, in-room cone-beam computed tomography (CBCT) assessed air gaps and repositioning was performed if necessary. IORT was delivered with low-energy X-rays (20-30 Gy on the applicator surface). Post-hoc Monte Carlo simulations quantified air gaps, reconstructed delivered dose distributions, and evaluated organ-at-risk (OAR) exposure. The primary endpoint was workflow feasibility and dosimetric integrity. Secondary endpoints included local control (LC) and radiation necrosis (RN). Results Between April 2023 and March 2024, 19 patients were enrolled and received image-guided IORT, which was successfully performed in all patients. The median preoperative BM volume (range) was 12.0 (3.8-70.5) cm3. The median IORT prescription dose was 30 (20-30) Gy, delivered with a spherical applicator of 25 (15-35) mm. Median air-gap volume was 0.12 (0.01-1.55) cm3. Three patients (16%) required applicator repositioning after CBCT, which reduced air gaps without clinically meaningful increases in OAR exposure. Median optic chiasm and brainstem D0.1 ccm were 0.31 (0.05-4.12) and 0.32 (0.05-2.82) Gy, respectively, whereas median brain V12 Gy was 0.67% (0.04-3.44%). Minimum dose within air pockets remained substantial (median Dmin 6.35 Gy), indicating limited attenuation. After a median follow-up of 23 (1-32) months, the 1-year LC rate was 88% and overall survival 52%. No RN attributable to IORT was observed. Conclusions CBCT verification demonstrates that small applicator-cavity air gaps are common but dosimetrically and clinically negligible when IORT is performed in an experienced setting. Monte Carlo reconstruction confirms robust dose delivery with excellent normal brain sparing for immediate high-dose single-fraction irradiation. The limited clinical impact of minor geometric imperfections may partly reflect the steep dose gradients and biological effectiveness of low-energy X-rays. These findings provide the first prospective validation of image-guided IORT and support its reliability as an adjuvant treatment for resectable BMs.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers