Helium pencil beam commissioning and beam modeling.
Lukas Martin, Barbara Knäusl, Peter Kuess, Dietmar Georg, Hugo Palmans, Lorenz Wolf, Nadia Gambino, Fabio Farinon,
MedAustron Ion Therapy Center, Marie Curie-Straße 5, Wiener Neustadt, A-2700, Austria. · RaySearch Laboratories AB, Eugeniavägen 18C, Stockholm, 113 68, Sweden.
Key points
From the paper's conclusion- Stable scanned beam delivery with helium ions was established. Validation demonstrated strong agreement between measurements, TPS calculations, and MC simulations, supporting research and future clinical application.
Abstract
BACKGROUND AND PURPOSE: Helium ions combine reduced lateral scattering compared to protons and a lower fragmentation tail than carbon ions, enabling sharp dose gradients and improved normal tissue sparing. This study reports the commissioning of a scanned helium pencil beam line and validation of its corresponding beam model. MATERIALS AND METHODS: Synchrotron-based helium ion beams were commissioned covering energies from 54.6 to 402.8 MeV/u. Depth-dose curves were measured and absolute dose calibration was performed. Beam optics (spot size, position, and intraspill stability) were evaluated for various spill lengths. A beam model was implemented in the RayStation treatment planning system (TPS) and validated through 2D absolute dose measurements in homogeneous and heterogeneous phantoms and 3D measurements of cubic spread-out Bragg peak fields. Gamma-index analysis and Monte Carlo (MC) simulations with GATE/Geant4 were performed for benchmarking. RESULTS: Measured ranges agreed with MC simulations within ± 0.3 mm. Spot sizes decreased with energy, independently of the spill length. Spot positions remained within ± 0.5 mm and intraspill variations were ≤ 0.2 mm (position) and ≤ 5.4 % (size). TPS-predicted doses agreed within 0.1 %. For 3D validations in homogeneous phantoms, the dose differences were generally within 2 %. Median gamma pass rates exceeded 90 % for 3 %/1.5 mm and 95 % for 5 %/1.5 mm. For the heterogeneous phantom, differences were within -3.2 %. CONCLUSIONS: Stable scanned beam delivery with helium ions was established. Validation demonstrated strong agreement between measurements, TPS calculations, and MC simulations, supporting research and future clinical application.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biophysical characterization of a helium beam: In-field and out-of-field microdosimetry and RBE.Medical physics · 2026
- Distance-dependent influence of scoring voxel size on Monte Carlo dosimetry of an iridium-192 HDR brachytherapy sourceEPJ Web of Conferences · 2026
- Comparison of experimental and radiobiological model-predicted survival of human laryngeal squamous cell carcinoma cells in a helium spread-out Bragg peak.Physics and imaging in radiation oncology · 2026
- SRS/SBRT commissioning process with a hybrid beam data model in new preconfigured CyberKnife system.Journal of applied clinical medical physics · 2026
- Comparison of proton, helium-ion, and carbon-ion CT imaging for hadron radiotherapy using Monte Carlo simulations.Physics in medicine and biology · 2026
- VHEE beam parameter extraction from depth-dose and lateral profile measurements: a Monte Carlo parametric study.Radiological physics and technology · 2026