Comparison of proton, helium-ion, and carbon-ion CT imaging for hadron radiotherapy using Monte Carlo simulations.
Zsófia Jólesz, Gábor Bíró, G Papp, Gergely Gábor Barnaföldi
HUN-REN Wigner Research Centre for Physics, Konkoly-Thege Miklós út 29-33, Budapest, 1121, Hungary. · Theoretical physics, Wigner RCP, Konkoly-Thege Miklós út 29-33, Budapest, 1121, Hungary.
Key points
From the paper's conclusion- To perform a comprehensive comparative analysis of proton, helium-ion, and carbon-ion computed tomography (CT) as direct imaging modalities for hadron therapy treatment planning, focusing on Relative Stopping Power (RSP) reconstruction accuracy and patient radiation dose.
Approach: High-fidelity Monte Carlo simulations were conducted using the GATE/Geant4 platform to model a standard CTP404 ph
Abstract
To perform a comprehensive comparative analysis of proton, helium-ion, and carbon-ion computed tomography (CT) as direct imaging modalities for hadron therapy treatment planning, focusing on Relative Stopping Power (RSP) reconstruction accuracy and patient radiation dose. Approach: High-fidelity Monte Carlo simulations were conducted using the GATE/Geant4 platform to model a standard CTP404 phantom. RSP maps were reconstructed using two iterative algorithms: the Simultaneous Algebraic Reconstruction Technique (SART) and the Richardson-Lucy (RL) deconvolution. Imaging performance was evaluated by comparing reconstructed RSP values against ground truth data for various tissue-equivalent inserts, while integral doses were estimated for a human head geometry using a Shepp-Logan-like phantom. Main results: All investigated particle modalities demonstrated a significant dose reduction compared to conventional X-ray CT protocols (which are approximately 40 mGy). The estimated imaging doses were 0.5 mGy for protons, 1.3 mGy for helium ions, and 7.5 mGy for carbon ions. In terms of accuracy, carbon-ion imaging achieved promising accuracy for soft-tissue materials (mean absolute error <0.5%). Helium ions offered a balanced performance with sub-1% errors for most materials and a dose burden significantly lower than carbon ions. Protons exhibited the widest range of RSP deviations. Significance: Direct particle imaging eliminates the systematic uncertainties inherent in photon-to-hadron conversion, which typically introduce RSP deviations of 1.6%-3.5% in standard single-energy X-ray CT. While carbon ions provide better RSP reconstruction precision essential for complex treatment plans in heterogeneous anatomy, helium and proton imaging offer exceptional dose sparing, making them particularly advantageous for pediatric patients and frequent adaptive replanning scenarios.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Proton treatment-plan dose verification with a DPM-based Monte Carlo engine.Frontiers in oncology · 2026
- Monte Carlo calculated beam quality correction factors for light ions used in particle therapyPhysics in Medicine and Biology · 2026
- Patient-Specific Stopping Power Calibration of Treatment Planning X-Ray CT Using Integrated Depth-Dose Profiles Acquired by Proton Transmission Imaging - A Feasibility Study.Physics in medicine and biology · 2026
- Monte Carlo analysis of depth dose distributions of 4He and 12C ion beams from 0.1 to 10 GeV in a human head phantom.Radiation and environmental biophysics · 2026
- LETd-dependent RBE conversion between LEM1- and mMKM-based dose systems and characterisation of LETd-dependent uncertainty in carbon ion radiotherapy.Physics in medicine and biology · 2026
- Image quality and artefact reduction for tantalum clips in ocular proton therapy using photon-counting CT.European radiology experimental · 2026