Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology

Beyond CBCT: real-time motion management in spine and non-spine bone SBRT using stereoscopic X-rays and surface guidance.

Adrián Gutiérrez, Kyra Dierinck, Tim Everaert, Selma Ben Mustapha, Guy Soete, Mark De Ridder, Thierry Gevaert

Department of Radiotherapy, UZ Brussel, Vrije Universiteit Brussel, Brussels, Belgium. Electronic address: adrian.gutierrez@uzbrussel.be. · Department of Radiotherapy, UZ Brussel, Vrije Universiteit Brussel, Brussels, Belgium.

Observational · n=71

Key points

From the paper's conclusion
  • Stereoscopic X-rays are a fast and efficient method to achieve sub-millimetre positioning accuracy equivalent to CBCT for spine and pelvis bone SBRT, effectively reducing setup residuals and intra-fraction errors and enabling the use of small margins. Tailored IGRT strategies are recommended to manage location-specific challenges like respiratory motion. Surface guidance functions as a complementary real-time alert of patient motion in between IGRT procedures.

Abstract

BACKGROUND: Stereotactic body radiotherapy (SBRT) for spine and non-spine bone metastases requires high precision due to sharp dose gradients and proximity to organs-of-interest. This study evaluates an image-guided radiotherapy (IGRT) workflow combining stereoscopic X-rays (XR) and surface guidance (SG) for setup, positioning, and intra-fraction monitoring. METHODS: We retrospectively analysed 92 bone metastases across 71 consecutive patients treated with SBRT (≤5 fractions). Patient setup and verification was performed with iterative XR and CBCT. Intra-fraction motion was monitored using stereoscopic gantry-triggered XR and continuous SG. Equivalence between XR and CBCT was evaluated using paired two one-sided tests. XR-SG correlations, sensitivity, and specificity of the monitoring system were analysed across anatomical regions. RESULTS: XR and CBCT showed equivalence for all spinal and pelvic targets. Iterative setup XR reduced mean 3D residuals from 1.8±2.1mm down to 0.3±0.2mm. During beam delivery, 95% of intra-fraction XR verifications for spine and pelvis locations remained within 1.5mm/1.5°. SG showed strong correlations with XR in the lateral (r = 0.872) and roll (r = 0.881) directions, but moderate to weak agreement in directions affected by respiratory motion. CONCLUSIONS: Stereoscopic X-rays are a fast and efficient method to achieve sub-millimetre positioning accuracy equivalent to CBCT for spine and pelvis bone SBRT, effectively reducing setup residuals and intra-fraction errors and enabling the use of small margins. Tailored IGRT strategies are recommended to manage location-specific challenges like respiratory motion. Surface guidance functions as a complementary real-time alert of patient motion in between IGRT procedures.

Read the paper