Impact of Radiation-Attenuating Drape on Exposure in a Contemporary Cardiac Catheterization Laboratory: The ATTENUATE Trial.
Giorgio A Medranda, Dimitrios Bliagos, Brian C Case
St Francis Heart Center at Good Samaritan University Hospital, West Islip, New York. · NYU Langone Hospital - Long Island, Mineola, New York.
Key points
From the paper's conclusion- In the largest randomized controlled trial to date evaluating the RADPAD protective scatter-radiation absorbing shield, encompassing contemporary coronary and structural CCL procedures, use of the protective scatter-radiation absorbing shield led to a marked decrease in proximal operator radiation exposure when adjusted for the total radiation delivered during each procedure.
Abstract
BACKGROUND: Despite recent improvements in radiation safety, interventionalists are increasingly exposed to radiation during cardiac catheterization laboratory (CCL) procedures. The RADPAD was designed as a protective scatter-radiation absorbing shield with early studies demonstrating a 20% to 62% reduction in scatter-radiation. The objective of this study was to examine the impact of the protective scatter-radiation absorbing shield in a large contemporary randomized controlled trial across multiple CCL procedures. METHODS: The investigator-initiated, prospective, randomized, controlled ATTENUATE (rAdpad proTecTion drapE iN redUcing rAdiaTion Exposure) trial randomized CCL procedures 1:1 to use of the RADPAD vs no use of the RADPAD. The primary outcome of interest was the most proximal operator's dose area product (DAP)-normalized operator dose (E). RESULTS: A total of 1000 CCL procedures were randomized 1:1 to use of the protective scatter-radiation absorbing shield (n = 500) vs no use (n = 500). The use of the protective scatter-radiation absorbing shield was associated with a 45.1% relative reduction in the DAP-normalized operator dose compared with no use of the protective scatter-radiation absorbing shield (0.64 vs 0.35 μSv/Gy⋅cm2; P < .0001). Coronary cases (n = 720) saw a 21.9% relative reduction in the DAP-normalized E using the RADPAD (0.33 vs 0.43 μSv/Gy⋅cm2; P < .001), and structural cases (n = 210) saw a 54.4% relative reduction in the DAP-normalized E using the RADPAD (0.30 vs 0.66 μSv/Gy⋅cm2; P = .1094). CONCLUSIONS: In the largest randomized controlled trial to date evaluating the RADPAD protective scatter-radiation absorbing shield, encompassing contemporary coronary and structural CCL procedures, use of the protective scatter-radiation absorbing shield led to a marked decrease in proximal operator radiation exposure when adjusted for the total radiation delivered during each procedure.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Randomized Evaluation of a Textile-Based Enhanced Radiation Protection Device Versus Control Shielding in the Cardiac Catheterization Laboratory.Journal of the Society for Cardiovascular Angiography & Interventions · 2026
- Evaluating Standard Shielding Versus Novel Enhanced Radiation Protection Devices in the Cardiac Catheterization Laboratory.Journal of the American Heart Association · 2026
- Clinical evaluation of radiation-shielding effects of a head-and-neck X-ray protective device during cardiovascular interventional procedures.Radiological physics and technology · 2026
- Real-time Radiation Exposure Monitoring in the Cardiac Catheterization Laboratory Using a Digital RRPD Dosimeter.Health physics · 2026
- Spatial distribution of X-ray attenuation in a full-face shield: a multipoint in vivo study in interventional cardiology.Radiation protection dosimetry · 2026
- Operator radiation exposure between left distal and right conventional radial access in patients undergoing PCIEuropean Heart Journal · 2026