Radiological physics and technology

Clinical evaluation of radiation-shielding effects of a head-and-neck X-ray protective device during cardiovascular interventional procedures.

Toshimitsu Sato, Akihiro Shinobu, Takanobu Hino, Chika Yamazaki, Yoichi Eguchi, Toshikazu Saida, Koichi Chida

Department of Radiology, Yamagata University Hospital, 2-2-2 Iida-nishi, Yamagata, 990-9585, Japan. tssato@med.id.yamagata-u.ac.jp. · Department of Radiology, Yamagata University Hospital, 2-2-2 Iida-nishi, Yamagata, 990-9585, Japan.

Observational

Key points

From the paper's conclusion
  • These findings suggest that TG may be a useful option for reducing occupational eye lens dose among physicians performing cardiovascular interventional procedures under routine clinical conditions.

Abstract

This study evaluated the radiation-shielding effects of a novel head-and-neck X-ray protective device, Triple Guard (TG; Maeda Co., Ltd, Tokyo, Japan), on occupational eye lens dose under routine clinical conditions in interventional cardiologists (ICs) performing coronary interventions and cardiac electrophysiologists (EPs) performing electrophysiological procedures. This observational study was conducted during routine clinical practice between December 2021 and March 2023 and included four physicians (two ICs and two EPs). A DOSIRIS eye lens dosimeter was positioned on the outer side of the left lens of the protective eyewear of each operator. The radiation-shielding effects of TG were evaluated in 289 procedures. Monthly DOSIRIS values obtained during the TG and non-TG periods were compared after normalization by the number of procedures, fluoroscopy time, air kerma at the patient entrance reference point (Ka,r), and air kerma-area product (PKA). For the ICs, the mean radiation-shielding effects were 66% per procedure, 74% per minute of fluoroscopy, 66% per unit Ka,r, and 68% per unit PKA. For the EPs, the corresponding values were 93%, 81%, 88%, and 89%, respectively. These findings suggest that TG may be a useful option for reducing occupational eye lens dose among physicians performing cardiovascular interventional procedures under routine clinical conditions. However, because this was a single-center observational study involving only four operators, further validation through multicenter studies and direct comparisons with conventional protective devices is warranted.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers