Performance-sensitivity analysis of atlantoaxial fixation constructs using finite element modelling.

Rahul Singh Dhari, Tiffani Tong, Jessica Gaff, Alana Celenza, Moreica Pabbruwe, Gregory Cunningham, Ian J Davies

School of Civil and Mechanical Engineering, Curtin University, Kent Street, Bentley, WA 6102, Australia. Electronic address: rahul.dhari@curtin.edu.au. · School of Civil and Mechanical Engineering, Curtin University, Kent Street, Bentley, WA 6102, Australia.

Computational

Key points

From the paper's conclusion
  • These findings demonstrate that atlantoaxial fixation performance is governed by a trade-off between absolute stability, stress demand, and sensitivity to cortical and cancellous bone stiffness.

Abstract

Atlantoaxial instability (AAI) is commonly treated using posterior fixation constructs that restrict relative C1-C2 motion and promote fusion. Although transarticular screw fixation, Goel-Harms screw-rod fixation, and combined fixation are established clinically, their relative biomechanical performance is often assessed primarily through displacement or range of motion reduction. The present study develops a three-dimensional finite element (FE) model of the C1-C2 segment and introduces a performance-sensitivity framework to evaluate fixation stability, implant stress demand, and robustness to bone-property variation. The model was benchmarked against published numerical and experimental data for intact and instrumented configurations. Flexion, extension, lateral bending, and axial rotation were simulated using a 1.5 N⋅m moment and a 40 N compressive preload. Cortical and cancellous elastic moduli were systematically varied to quantify displacement sensitivity, stress sensitivity, and an overall performance-sensitivity score for each construct. At the reference material condition, the combined construct provided the greatest reduction in relative C1-C2 displacement and generally produced the lowest peak construct stresses. However, the sensitivity analysis showed that the most stable construct was not necessarily the least sensitive to bone-property variation. The Goel-Harms construct exhibited lower material sensitivity in several loading cases, but this occurred at higher absolute displacement and stress levels. Increasing cancellous stiffness generally reduced displacement and construct stress, whereas increasing cortical stiffness reduced displacement but often increased construct stress, indicating increased implant stress demand. These findings demonstrate that atlantoaxial fixation performance is governed by a trade-off between absolute stability, stress demand, and sensitivity to cortical and cancellous bone stiffness. The proposed framework provides a mechanics-based basis for interpreting fixation behaviour beyond conventional displacement-based comparison.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers