Biomechanical Comparison of Three Fixation Constructs for Tile Type C1.2 Pelvic Ring Fractures: A Finite Element Analysis
Adrian Claudiu Carp, Bogdan Veliceasa, Dmour Awad, Alexandru Filip, Mihaela Perțea, Norin Forna, Bogdan Puha, Ștefan Dragoș Tîrnovanu,
Grigore T. Popa University of Medicine and Pharmacy · Gheorghe Asachi Technical University of Iași
Key points
From the paper's conclusion- The aim of this study was to compare, by means of finite element analysis, the biomechanical performance of three different methods of osteosynthesis for Tile type C1.2 pelvic ring fractures: a transiliac plate, one iliosacral screw and two anterior reconstruction plates on the sacroiliac joint.
Abstract
Fractures of the pelvic ring are among the most severe injuries in orthopaedic practice and Tile type C lesions are characterized by complete disruption of the posterior arch with both vertical and rotational instability. The optimal construct for posterior ring fixation remains a matter of debate. The aim of this study was to compare, by means of finite element analysis, the biomechanical performance of three different methods of osteosynthesis for Tile type C1.2 pelvic ring fractures: a transiliac plate, one iliosacral screw and two anterior reconstruction plates on the sacroiliac joint. A three-dimensional model of an intact pelvis was reconstructed from computed tomography images of a healthy adult male. A Tile type C1.2 injury pattern was created virtually, and three fixation constructs were designed in Ansys SpaceClaim according to manufacturer specifications. All materials were assumed to be homogeneous, isotropic and linearly elastic. Vertical loads of 400 N and 800 N were applied to the sacral endplate to simulate partial and full weight bearing, while the acetabular regions were constrained to represent standing stance. In this study, mechanical stability was operationally defined as resistance to global displacement under applied vertical load, with lower displacement indicating higher construct stiffness. Construct stiffness, total deformation and von Mises stress were assessed for bone and implants. For both loading conditions, the iliosacral screw construct showed the lowest overall displacement and provided the greatest stiffness. The transiliac plate construct presented larger displacements, whereas the anterior reconstruction plate construct provided intermediate stability with higher stresses at the sacroiliac joint. Among the analyzed constructs, the iliosacral screw provided the greatest stiffness and lowest overall displacement, suggesting superior mechanical performance under vertical loading conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of novel anterior fixation method, W-plate, versus superior parasymphyseal plating in anterior posterior compression type III pelvic ring injuries: a biomechanical study.Journal of orthopaedic trauma · 2026
- Vertically Unstable Pelvic Ring Injuries: Impact of Multilevel and Transsacral Fixation Strategy on Failure Rate.Journal of orthopaedic trauma · 2026
- Staged Reduction and Combined Anterior and Posterior Minimally Invasive Fixation for a Highly Unstable Open-Book Pelvic Ring Fracture: A Case Report.Cureus · 2026
- Fracture location-type synergy impacts stability in surgical stabilization of rib fractures.Journal of biomechanics · 2026
- Axial Stiffness Advantage of Nail–Plate Fixation in 101 Patient-Specific Femora: A Finite Element StudyJournal of Clinical Medicine · 2026
- Biomechanical comparison of multirod constructs for severe thoracolumbar Pott's deformity following posterior vertebral column resection: a finite element study.Asian spine journal · 2026