Biomechanical comparison of seven occipital-cervical fixation constructs using finite element modeling.
Ian Polyzois, Jaskaran Singh, Sara Gustafson, Trevor Gascoyne, Franck Le Navéaux, Pierre Legay, Michael Goytan
Orthopaedic Innovation Centre, Winnipeg, MB, Canada. · Max Rady College of Medicine, University of Manitoba, Winnipeg, MB, Canada.
Key points
From the paper's conclusion- A construct composed of medially fixed one-hole and two-hole clamps provided the best balance of bending stiffness and minimized occipital screw pullout reaction forces. Aligning screws near the median plane, using bilateral clamp fixation, and avoiding components that mechanically join the two rods improved construct performance. Increasing occipital screw diameter from 4.5 mm to 5.0 mm had little biomechanical impact. Stress localization at occipital component-rod interfaces suggests a potential weak point with risk of notching and fatigue failure.
Abstract
STUDY DESIGN: This was a basic science study. OBJECTIVE: The objective of this study was to determine which occipital-cervical (OC) fixation construct best balances construct stiffness with minimization of occipital screw pullout forces. MATERIALS AND METHODS: Finite element analysis was used to simulate seven OC construct variations spanning across C1 with fixation in C2 and C3. Variations included the use of occiput rods, keel plates with varying width, two-hole clamps, two-hole clamps in addition to medial or lateral one-hole clamps, and occiput screw diameter. Modeling was in accordance with ASTM F2706/ISO 12189 guidelines. Materials were in accordance with ASTM F1295/ASTM F1839/ISO 10243 guidelines. Outcomes included bending stiffness, axial screw pullout reaction forces, and von Mises stress distribution to identify stress localization and predicted failure modes. Constructs were ranked using multiattribute utility theory (MAUT). RESULTS: Construct E (comprised two-hole clamps with added one-hole medial clamps) demonstrated the highest bending stiffness (820 N.mm/deg) and the highest MAUT utility score (0.79), representing the optimal balance of stiffness and minimized screw pullout forces. Construct F (comprised two-hole clamps alone) demonstrated the lowest bending stiffness (509 N.mm/deg). CONCLUSIONS: A construct composed of medially fixed one-hole and two-hole clamps provided the best balance of bending stiffness and minimized occipital screw pullout reaction forces. Aligning screws near the median plane, using bilateral clamp fixation, and avoiding components that mechanically join the two rods improved construct performance. Increasing occipital screw diameter from 4.5 mm to 5.0 mm had little biomechanical impact. Stress localization at occipital component-rod interfaces suggests a potential weak point with risk of notching and fatigue failure.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance-sensitivity analysis of atlantoaxial fixation constructs using finite element modelling.Journal of the mechanical behavior of biomedical materials · 2026
- Biomechanical Comparison of Composite versus Screw-Rod Fixation for Posterior C5-C6 Fusion.Global spine journal · 2026
- Biomechanical comparison of multirod constructs for severe thoracolumbar Pott's deformity following posterior vertebral column resection: a finite element study.Asian spine journal · 2026
- Biomechanical study of posterior lumbar fixation combined with partial laminectomy using finite element analysisBIO Web of Conferences · 2026
- Feasibility of Occipital Condyle Screw Fixation for Craniocervical Instability: Integrating Morphometric Evidence, Anatomical Parameters, and Experience From Surgical Series.Neurospine · 2026
- Biomechanical advantages of double-trajectory screws technique combined with multiple-rod constructs in lumbar pedicle subtraction osteotomy: a finite element analysis study.Frontiers in bioengineering and biotechnology · 2026