OPTIMISING ACCURACY OF COMPUTER-ASSISTED IMPLANT SURGERY USING NOVEL DRILLING KEY DESIGN: SIMULATION-BASED STUDY.
Thaysan Kaing, Sirida Arunjaroensuk, Xin Hui Yeo, Min Thu Kha Ko Ko, Nikos Mattheos, Atiphan Pimkhaokham
Oral and Maxillofacial surgery and Digital Implant Surgery Research Unit, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand; Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand. · Oral and Maxillofacial surgery and Digital Implant Surgery Research Unit, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand; Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand; Department of Dental Medicine, Karolinska Institute, Stockholm, Sweden.
Key points
From the paper's conclusion- The findings imply that the novel key design provides a targeted advantage in reducing angular deviation in both bone densities while linear deviations at the platform and apex remain influenced by bone density and other potential sources of error within the guided workflow.
Abstract
BACKGROUND: Accurate implant positioning remains the goal of guided surgery for improving surgical outcomes and minimizing complications. The aim is to compare the accuracy of standard versus novel drilling key in static-dynamic computer assisted implant surgery across D2 and D3 bone densities relative to a zero-deviation benchmark. METHODS: Sixty-four bone-level tapered implants were placed in 32 models across four groups (n = 16), stratified by drilling key designs (standard vs novel) and bone density (D2 vs D3). Preoperative cone-beam computed tomography (CBCT) was imported into the navigation system for implant planning. Surgical guide was designed to match the navigation planning. Implant placement followed the manufacturer protocols. Postoperative CBCT was superimposed on pre-operative plan to measure 3D platform, 3D apex, and angular deviations. Two-way ANOVA, Levene, Kruskal-Wallis, and post-hoc Mann-Whitney U tests were used (p < 0.05). RESULTS: The novel drilling key significantly reduced angular deviation compared to the standard key (1.84°± 1.58 vs 0.21°± 0.56, p < 0.001); 3D platform (0.78 ± 0.30 vs 0.74 ± 0.17 mm, p = 0.262) and apex (0.82 ± 0.31 vs 0.93 ± 0.41 mm, p = 0.596) were comparable. D3 bone showed higher 3D apex (p = 0.017) and angular deviations (p = 0.044) than D2. CONCLUSION: The findings imply that the novel key design provides a targeted advantage in reducing angular deviation in both bone densities while linear deviations at the platform and apex remain influenced by bone density and other potential sources of error within the guided workflow. CLINICAL SIGNIFICANCE: This study addresses accuracy in implant dentistry in technically challenging cases such as long free drilling distance where low sleeve positioning is not feasible.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Short implant placement using dynamic and robotic computer-assisted implant surgery (d-CAIS and r-CAIS): a controlled in vitro study.Journal of dentistry · 2026
- Accuracy, insertion torque and ISQ of keyless versus drill-key guided implant surgery: in vitro comparison of two implant macro-designs in healed ridges and molar extraction sites.Journal of dentistry · 2026
- Influence of clinician experience with a static computer-assisted implant surgery workflow on implant-placement accuracy across three surgical-guide designs: a clinical trial.Journal of dentistry · 2026
- Impact of Level of Guidance on the Accuracy of Static Computer-Assisted Implant Surgery (s-CAIS): An In Vitro Study.Clinical oral implants research · 2026
- Accuracy of Static Computer-Assisted Immediate and Delayed Anterior Implant Placement by Novice Clinicians.European journal of dentistry · 2026
- Accuracy of Computer-Guided Dental Implant Placement: A Clinical Comparison of Three Surgical Guide TypesJournal of Clinical Medicine · 2025