Extraoral relocation of implant scanbody geometries: An in vitro comparison of two workflows for complete-arch implant position transfer.

Gennaro Ruggiero, Roberto Sorrentino, Francesco Guido Mangano, Jaafar Mouhyi, Fernando Zarone

Department of Neurosciences, Reproductive and Odontostomatological Sciences, Division of Prosthodontics, Scientific Unit of Digital Dentistry, University "Federico II" of Naples, Naples, Italy. Electronic address: gennaro_ruggiero@hotmail.it. · Department of Neurosciences, Reproductive and Odontostomatological Sciences, Division of Prosthodontics, Scientific Unit of Digital Dentistry, University "Federico II" of Naples, Naples, Italy.

Preclinical

Key points

From the paper's conclusion
  • Under the tested in vitro conditions, the integrated GXE workflow showed lower global surface deviation and smaller detrended translational and angular deviations than conventional ISBs. These metrological findings provide a basis for further prosthetic-fit and clinical validation.

Abstract

OBJECTIVES: To compare the metrological performance of an integrated extraoral GeoXact scanbody workflow (GXE; GeoXact® extraoral kit, Biotec srl, Vicenza, Italy) with a conventional intraoral implant scanbody workflow (ISB; Osstem TSIII, 4.8-mm platform; Osstem Implant Ltd., Busan, Republic of Korea) for complete-arch implant position transfer. METHODS: A maxillary edentulous cast containing 6 digital laboratory analogs (TSMDLA, Osstem Implant Ltd., Busan, Republic of Korea) was scanned 15 times per workflow with one intraoral scanner (Primescan 2; Sirona Dental, Bensheim, Germany). Each experimental scan was compared with a workflow-specific structured-light reference after CAD-library replacement. The prespecified primary outcome was scan-averaged detrended translational magnitude; key secondary outcomes were global mean surface deviation and scan-averaged detrended angular deviation. Site-level and interimplant analyses were exploratory. RESULTS: GXE showed lower detrended translational magnitude (96 ± 52 µm vs 210 ± 72 µm; P < .001), global mean surface deviation (8 ± 3 µm vs 18 ± 16 µm; P = .041), and detrended angular deviation (0.121 ± 0.059° vs 0.320 ± 0.045°; P < .001) than conventional ISBs. CONCLUSIONS: Under the tested in vitro conditions, the integrated GXE workflow showed lower global surface deviation and smaller detrended translational and angular deviations than conventional ISBs. These metrological findings provide a basis for further prosthetic-fit and clinical validation. CLINICAL SIGNIFICANCE: Under controlled in vitro conditions, the GXE workflow yielded lower global surface deviation and smaller relative translational and angular discrepancies than conventional ISBs.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers