Extraoral relocation of implant scanbody geometries: An in vitro comparison of two workflows for complete-arch implant position transfer.
Gennaro Ruggiero, Roberto Sorrentino, Francesco Guido Mangano, Jaafar Mouhyi, Fernando Zarone
Department of Neurosciences, Reproductive and Odontostomatological Sciences, Division of Prosthodontics, Scientific Unit of Digital Dentistry, University "Federico II" of Naples, Naples, Italy. Electronic address: gennaro_ruggiero@hotmail.it. · Department of Neurosciences, Reproductive and Odontostomatological Sciences, Division of Prosthodontics, Scientific Unit of Digital Dentistry, University "Federico II" of Naples, Naples, Italy.
Key points
From the paper's conclusion- Under the tested in vitro conditions, the integrated GXE workflow showed lower global surface deviation and smaller detrended translational and angular deviations than conventional ISBs. These metrological findings provide a basis for further prosthetic-fit and clinical validation.
Abstract
OBJECTIVES: To compare the metrological performance of an integrated extraoral GeoXact scanbody workflow (GXE; GeoXact® extraoral kit, Biotec srl, Vicenza, Italy) with a conventional intraoral implant scanbody workflow (ISB; Osstem TSIII, 4.8-mm platform; Osstem Implant Ltd., Busan, Republic of Korea) for complete-arch implant position transfer. METHODS: A maxillary edentulous cast containing 6 digital laboratory analogs (TSMDLA, Osstem Implant Ltd., Busan, Republic of Korea) was scanned 15 times per workflow with one intraoral scanner (Primescan 2; Sirona Dental, Bensheim, Germany). Each experimental scan was compared with a workflow-specific structured-light reference after CAD-library replacement. The prespecified primary outcome was scan-averaged detrended translational magnitude; key secondary outcomes were global mean surface deviation and scan-averaged detrended angular deviation. Site-level and interimplant analyses were exploratory. RESULTS: GXE showed lower detrended translational magnitude (96 ± 52 µm vs 210 ± 72 µm; P < .001), global mean surface deviation (8 ± 3 µm vs 18 ± 16 µm; P = .041), and detrended angular deviation (0.121 ± 0.059° vs 0.320 ± 0.045°; P < .001) than conventional ISBs. CONCLUSIONS: Under the tested in vitro conditions, the integrated GXE workflow showed lower global surface deviation and smaller detrended translational and angular deviations than conventional ISBs. These metrological findings provide a basis for further prosthetic-fit and clinical validation. CLINICAL SIGNIFICANCE: Under controlled in vitro conditions, the GXE workflow yielded lower global surface deviation and smaller relative translational and angular discrepancies than conventional ISBs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of Scan Body Geometry and CAD Software on the Accuracy of Implant-Position Transfer: An In Vitro Study.Bioengineering (Basel, Switzerland) · 2026
- A centripetally rotating implant scan body concept for complete-arch scanning: An in vitro studyJournal of Dentistry · 2026
- Trueness and Precision of Three Novel Scan Body Designs with Horizontal Extensions for Complete-Arch Implant Impressions: A Comparative In Vitro Study with Intraoral Scanning and Photogrammetry Systems.Journal of functional biomaterials · 2026
- Accuracy of virtual patient reconstruction for complete arch implant-supported prostheses with extraoral scan body-assisted registration.The Journal of prosthetic dentistry · 2026
- Implant Position Accuracy of a Dual-Function Bite Holder for Single-Visit Registration of Complete-Arch Implants and Maxillomandibular Relationship: An In Vitro Study.The International journal of oral & maxillofacial implants · 2026
- Linear and Angular Deviations of Complete-Arch Digital Implant Impressions Using Three Scan Body Systems: An In Vitro Study.Dentistry journal · 2026