Implant Position Accuracy of a Dual-Function Bite Holder for Single-Visit Registration of Complete-Arch Implants and Maxillomandibular Relationship: An In Vitro Study.

Ra'fat I Farah, Abdulmajeed Alwably, Abdulaziz Alfurayhi, Asim Alfawzan

Preclinical · n=10

Key points

From the paper's conclusion
  • Digital substitution of bite holders with scan bodies achieved clinically acceptable accuracy for implant position registration under in vitro conditions, with differences from direct scan body scanning within the predefined equivalence margins This design enables simultaneous single-visit capture of implant positions and maxillomandibular relationships (MMR), warranting clinical validation.

Abstract

PURPOSE: To evaluate the accuracy of implant spatial positions obtained through a proposed bite holder digital substitution workflow compared to conventional direct scan body scanning. MATERIAL AND METHODS: A mandibular model with four implant analogs was fabricated and digitized with a desktop scanner to establish the reference standard. Two protocols were compared: direct scan body scanning (n=10) and bite holder scanning with digital substitution (n=10), both performed with an intraoral scanner. In the substitution method, bite holders were virtually replaced with scan bodies using a superimposition workflow in computer-aided design (CAD) software. Three-dimensional RMS deviation, linear deviation, and angular deviation were quantified using metrology software. Hotelling's T² test compared overall accuracy between methods (α=0.05), followed by protected univariate tests with Bonferroni correction. An equivalence analysis was performed using the two one-sided tests procedure with clinical margins of ±100 μm and ±0.4°. RESULTS: Direct scanning achieved RMS, linear, and angular deviations of 56.3±14.5 μm, 67.5±15.7 μm, and 0.37±0.06°, respectively. The substitution method demonstrated 85.2±18.0 μm, 96.6±15.0 μm, and 0.63±0.07°, respectively. Direct scanning showed statistically superior accuracy (P<.001); however, the 90% confidence intervals of all mean differences fell entirely within the predefined equivalence margins, indicating that the differences were smaller than the clinically relevant limits. CONCLUSIONS: Digital substitution of bite holders with scan bodies achieved clinically acceptable accuracy for implant position registration under in vitro conditions, with differences from direct scan body scanning within the predefined equivalence margins This design enables simultaneous single-visit capture of implant positions and maxillomandibular relationships (MMR), warranting clinical validation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Implant Position Accuracy of a Dual-Function Bite Holder for Single-Visit Registration of Complete-Arch Implants and Maxillomandibular Relationship: An In Vitro Study — Science Skim