B-002 Cofactors in Focus: Pyridoxal 5'-phosphate Versus Non-Pyridoxal 5'-phosphate Alanine Aminotransferase and Aspartate Aminotransferase Assays

Woon Ting Priscilia Cheong, Sau Yeng Jayme Wong, Huan Thin Lee, Ying Choo, Gin Xiang Johnston Tang, Chin Pin Yeo

Key points

From the paper's conclusion
  • Our study evaluated the analytical performance of P5P-activated ALT and AST assays on Beckman Coulter AU5832 and DxC700AU clinical chemistry analysers and compared them with the existing non-P5P methods currently in routine use in our tertiary care hospital laboratory.

Abstract

Background Alanine aminotransferase (ALT) and aspartate aminotransferase (AST) are key biomarkers of hepatocellular injury. Conventional assays rely on endogenous pyridoxal 5'-phosphate (P5P) as a co-factor, the availability of which may vary among patients and lead to underestimation of enzyme activity, particularly in individuals with vitamin B6 deficiency. P5P-supplemented assays provide a standardised cofactor concentration, potentially improving measurement accuracy, method harmonisation and clinical reliability. Our study evaluated the analytical performance of P5P-activated ALT and AST assays on Beckman Coulter AU5832 and DxC700AU clinical chemistry analysers and compared them with the existing non-P5P methods currently in routine use in our tertiary care hospital laboratory. Methods The analytical evaluation of the P5P ALT and AST assays on the Beckman AU systems included assessments of imprecision, linearity, limit of detection (LoD), and carryover using our laboratory's technical performance protocols which are aligned with CLSI guidelines. Method comparison studies were performed using 100 residual serum specimens obtained from our hospital inpatients and outpatients with concentrations across the full analytical measuring interval of both P5P and non-P5P assays. Statistical analyses were performed using Analyze-It version 6.15. Results Study results demonstrated that imprecision, analytical sensitivity, and carryover for the Beckman P5P-activated ALT and AST assays were within specifications. Within-run and total imprecision CVs were < 3% for AU5832 and DxC 700AU. Linearity was verified across the analytical measuring range. LoD values ranged from 1.17–3.68 U/L. Carryover studies showed no significant carryover. Method comparison studies demonstrated consistently higher enzyme activities with P5P-activated assays compared to non-P5P assays. For ALT, Bland-Altman analysis showed a mean positive bias of +15.36% (95% CI: 13.35%–17.38%) on AU5832 and +14.97% (95% CI: 13.21%–16.72%) on DxC700AU. Passing–Bablok regression analysis demonstrated regression equations of [P5P ALT] = 1.14 × [non-P5P ALT] - 0.22 (R² = 0.99) for AU5832 and [P5P ALT] = 1.14 × [non-P5P ALT] - 0.29 (R² = 0.99) for DxC700AU. For AST, Bland–Altman analysis showed a consistent positive bias of +8.87% (95% CI: 6.08%–11.66%) on AU5832 and +8.94% (95% CI: 6.11%–11.76%) on DxC700AU. Passing–Bablok regression analysis demonstrated regression equations of [P5P AST] = 1.04 × [non-P5P AST] + 0.24 (R² = 0.99) for AU5832 and [P5P AST] = 1.04 × [non-P5P AST] + 1.31 (R² = 0.99) for DxC700AU. Conclusion The P5P-activated ALT and AST assays on the Beckman Coulter AU5832 and DxC700AU analysers met all verification criteria for precision, and carryover. Method comparison demonstrated consistent positive proportional bias for the P5P-activated assays, reflecting the effect of cofactor supplementation on measured enzyme activity. As the comparison utilised residual clinical specimens without stratification by vitamin B6 status, the specific contribution of P5P deficiency could not be directly quantified. Nevertheless, the observed systematic increase supports the presence of variable endogenous cofactor availability within the tested population. Adoption of P5P-activated assays is expected to reduce cofactor-related variability and improve assay standardisation, accuracy, and clinical reliability. Based on these findings, the laboratory will adopt P5P-activated ALT and AST assays to improve analytical harmonisation and support more accurate and reliable clinical interpretation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers