Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine

An inversion-forward method for γ-ray spectrum background subtraction based on the ML-EM inversion algorithm.

Wei Chen, Xuemei Wang, Wenbiao Liu, Lihao Wang, Zhihui Wang

Northwest Institute of Nuclear Technology, Xi'an, 710024, China. Electronic address: chwei2601@163.com. · Northwest Institute of Nuclear Technology, Xi'an, 710024, China.

Computational

Key points

From the paper's conclusion
  • To address the limitation of traditional methods in accurately subtracting backgrounds from complex γ-ray spectra, particularly those containing overlapping peaks, this paper proposes an inversion-forward method for γ-ray spectrum background subtraction based on the Maximum Likelihood-Expectation Maximization (ML-EM) inversion algorithm.

Abstract

γ-ray spectrum analysis is one of the widely used techniques in the field of nuclear radiation detection. Background subtraction from γ-ray spectra can improve the quantitative accuracy of γ-emitting radionuclides. To address the limitation of traditional methods in accurately subtracting backgrounds from complex γ-ray spectra, particularly those containing overlapping peaks, this paper proposes an inversion-forward method for γ-ray spectrum background subtraction based on the Maximum Likelihood-Expectation Maximization (ML-EM) inversion algorithm. A 4″ × 4″×16″ thallium-doped sodium iodide (NaI(Tl)) detector was used to acquire γ-ray spectra containing information from 241Am, 137Cs, 60Co, 152Eu, 133Ba, and 22Na. Focusing on 241Am, 137Cs, and 60Co while treating the other nuclides as interference sources, the proposed method was applied for background subtraction. The results were compared with those obtained using the Maestro32 software and the Statistics-sensitive Nonlinear Iterative Peak-clipping (SNIP) method. The results demonstrate that the background subtraction accuracy of the proposed method is significantly superior to the other two methods. After background subtraction, the maximum absolute deviation of the calculated activities of 241Am, 137Cs, and 60Co (where the 241Am characteristic peak overlaps with that of 133Ba, and the 60Co characteristic peaks overlap with those of 152Eu, 133Ba, and 22Na) from the reference values is only 5.6%, compared to 48.8% for Maestro32 and 72.5% for the SNIP method.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers