Site-specific field comparison of in situ NaI(Tl) gamma-ray spectrometry for ambient dose equivalent rate estimation and 137Cs spectral screening.
Hyungwoo You, Wonseok Bak, Gyoungjun Choi, Myeongsik Shin, Jun Yoo, Min-Kun Kim, Jangoh Kim, Ku Kang
CBRN Defense Research Institute, 06796 Seoul, Republic of Korea; Department of Radiological Science, Inje University, 50834 Gimhae-si, Republic of Korea. · Korea Atomic Energy Research Institute, 34057 Daejeon, Republic of Korea.
Key points
From the paper's conclusion- These results indicate that a G(E)-derived dose rate restricted to the terrestrial photon spectrum must be combined with a site-specific cosmic-ray estimate before it is compared with a survey-meter reading, and that a single, documented peak-extraction procedure is a prerequisite for spatial comparison of 137Cs in low-level NaI(Tl) spectra.
Abstract
In situ NaI(Tl) gamma-ray spectrometry is widely used in environmental radioactivity surveys. G(E)-function-based spectrum-dose conversion provides a practical means of deriving ambient dose equivalent rates from measured pulse-height spectra. However, the performance of G(E)-derived estimates relative to independently calibrated field instruments has seldom been quantified under realistic ground-source conditions. We compared G(E)-derived ambient dose equivalent rates from a portable NaI(Tl) spectrometer with simultaneous readings from a calibrated FH40G survey meter at ten co-located points at an agricultural site about 680 m above sea level on Jeju Island, Republic of Korea. The G(E)-derived rates were lower than the FH40G readings at every one of the ten points: the mean G(E)/FH40G ratio was 0.493 (Wilcoxon signed-rank W=0, p=0.002; Cohen's dz=-5.1) and the mean difference was -0.049μSvh-1 (95% CI -0.056 to -0.042μSvh-1). About 70% of this difference is accounted for by the cosmic-ray ionising component at the site elevation (35 nSv h-1), which the FH40G registers but which lies outside the 50 keV-3.1 MeV range of the G(E) summation. With this component added, the mean ratio is 0.863 ± 0.091. The remaining difference of 14% is systematic but within the expanded uncertainty of the comparison, provided that the survey meter's cosmic response is at least 0.74 of nominal. The same spectra were screened for 137Cs with one spectral-fitting procedure, applied identically to all ten spectra, that resolves the 662 keV peak from the adjacent 214Bi 609 keV line. A 137Cs photopeak was present at every point, at levels within a factor of two of each other (fitted areas 500-961 counts per 1198 s live time), consistent with residual global fallout, and no point showed a localised enhancement; net counts from regions of interest set manually at each point had suggested such an enhancement. These results indicate that a G(E)-derived dose rate restricted to the terrestrial photon spectrum must be combined with a site-specific cosmic-ray estimate before it is compared with a survey-meter reading, and that a single, documented peak-extraction procedure is a prerequisite for spatial comparison of 137Cs in low-level NaI(Tl) spectra. These findings characterise one agricultural site under a single detector and conversion-function configuration and are not presented as a general calibration result for NaI(Tl) spectrometry.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Environmental radiological baseline characterization of a proposed medical cyclotron facility in Indonesia using dual-detector measurements and geostatistical analysis.Environmental geochemistry and health · 2026
- Segmented G(E) Function Compensation Method for Gamma Dose Rate Meters Considering Asymmetric Error Distribution Across Energy RegionsSymmetry · 2026
- Urban Dose Rate Estimation Using Distributed Static D3S Gamma-Ray Spectrometers.Sensors (Basel, Switzerland) · 2026
- National intercomparison of environmental gamma radiation dose rate in China based on ice surface cosmic ray response measurements.Journal of environmental radioactivity · 2026
- Gamma-ray spectrum unfolding and air-absorbed dose-rate estimation using a two-stage spectral learning framework.Radiation protection dosimetry · 2026
- Operator-induced reduction of air kerma during backpack-based gamma-ray measurement.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026