ROC-based evaluation of tight gamma criteria for detecting MLC and jaw positioning errors in brain SRS using SRS MapCHECK.

Hyojun Park, Hyejo Ryu, Do Hoon Oh, Jin Hwa Choi, Minsoo Chun

Department of Radiation Oncology, Chung-Ang University Gwangmyeong Hospital, Gyeonggi-do 14353, the Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 03080, the Republic of Korea. · Department of Radiation Oncology, Chung-Ang University Gwangmyeong Hospital, Gyeonggi-do 14353, the Republic of Korea.

Key points

From the paper's conclusion
  • For Elekta Versa HD brain SRS QA using the SRS MapCHECK system, the 3%/0.5 mm criterion may be considered for institution-specific use, particularly for aperture-changing errors, after local validation.

Abstract

PURPOSE: Precise patient-specific quality assurance (QA) is critical for brain stereotactic radiosurgery (SRS). This study evaluated tight gamma criteria using the SRS MapCHECK high-resolution dual-layer detector. METHODS: Seventeen volumetric modulated arc therapy (VMAT) plans delivered on an Elekta Versa HD were measured under error-free conditions and after intentional 0.5-mm multi-leaf collimator (MLC) or 1.0-mm Y-jaw shifts. Gamma passing rates (GPRs) were calculated using nine criteria at low-dose thresholds of 10%, 20%, and 30%. Pooled and error-type-stratified receiver operating characteristic (ROC) analyses were performed, and confidence intervals for the area under the curve (AUC) were estimated using patient-level cluster bootstrap resampling. RESULTS: The conventional 3%/2 mm criterion showed limited pooled discrimination (AUC < 0.65). The 1%/0.5 mm criterion yielded the numerically highest pooled AUC of 0.830, whereas the 3%/0.5 mm criterion yielded an AUC of 0.796, with sensitivity and specificity of 0.765 at a cutoff of 88.8%. With the 3%/0.5 mm criterion, the AUCs were 0.872 for MLC aperture-changing errors, 0.581 for MLC translations, 0.871 for jaw aperture-changing errors, and 0.861 for jaw translations. MLC translations produced only minor dose-volumetric changes, whereas aperture-changing errors had larger effects on target coverage and normal-tissue dose. CONCLUSIONS: For Elekta Versa HD brain SRS QA using the SRS MapCHECK system, the 3%/0.5 mm criterion may be considered for institution-specific use, particularly for aperture-changing errors, after local validation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers