Statistical process control and error curve modelling identify superior pre-treatment verification metrics for nasopharyngeal carcinoma and cervical cancer VMAT: a retrospective study
Wenli Lu, Hanyin Zhang, Haixia Cui, Wei Huang, Qingfeng Jiang, Furong Lv, Xin Yi
Key points
From the paper's conclusion- We integrated statistical process control (SPC) with error curve modelling (ECM) to identify metrics detecting clinically significant errors, derive tolerance limits, and quantify undetected error ranges.
Abstract
Background and purpose Gamma analysis is widely used for volumetric modulated arc therapy (VMAT) verification, but its clinical relevance and sensitivity to delivery errors remain limited. We integrated statistical process control (SPC) with error curve modelling (ECM) to identify metrics detecting clinically significant errors, derive tolerance limits, and quantify undetected error ranges. Materials and methods This retrospective study included 120 VMAT plans (68 nasopharyngeal carcinoma [NPC], 52 uterine cervical cancer [UCC]). Five mechanical error types were introduced, yielding 1831 datasets. Under three global gamma criteria (3%/2 mm, 2%/2 mm, 1%/1 mm), gamma passing rates (GPRs) and mean gamma index (μGI) were computed in multiple regions. Metrics were screened by Spearman correlation, tolerance limits derived by SPC, undetected ranges quantified by ECM, and discrimination assessed at 3%/2 mm by ROC analysis in an independent patient-level validation set (628 datasets, 55 patients), against a clinical (|DE%PTV| ≥ 3%) and a mechanical (any error) standard. Results GPRswhole, μGIwhole, GPRs90% and μGI90% correlated strongly with PTV dose error (|r| ≥ 0.5). SPC on the development cohort gave 3%/2 mm lower limits for GPRswhole of 0.509 (NPC) and 0.470 (UCC) and upper limits for μGIwhole of 1.016 and 1.087; 97.4–100% of error-free validation plans fell within them. ECM, interpreted only where R² > 0.8, responded to MU and PTV dose errors, whereas random MLC, gantry and most systematic MLC curves were too poorly constrained to yield a detection limit. Against the clinical standard, AUCs were 0.999 (NPC) and 0.874 (UCC) for GPRswhole and 0.997 and 0.841 for μGIwhole, versus 0.605 and 0.810 for GPRs10%; in UCC the 90%-region metrics reached 0.940 and 0.947. At 90% specificity, sensitivities were 0.98–1.00 (NPC) and 0.70–0.83 (UCC) versus 0.37 and 0.58. Under the mechanical standard the advantage was smaller in NPC and absent in UCC. Conclusions Modulation-based metrics on clinically relevant regions identified clinically relevant dose deviations better than GPRs10%, which remained more sensitive to the mere presence of a perturbation; the two are complementary. These limits and undetected error ranges are site-specific and hypothesis-generating, requiring multi-centre validation before clinical acceptance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assessing Gamma Index Sensitivity to Selected Linear Accelerator Mechanical Errors Using Various Dosimetric Systems.Journal of clinical medicine · 2026
- Impact of daily machine performance check on prediction models for patient‐specific quality assurance of multiple‐target stereotactic radiosurgeryMedical Physics · 2026
- Sensitivity in error detection of an electronic portal imaging device for patient-specific QA in head and neck VMAT plans.Radiological physics and technology · 2026
- Machine learning-based prediction of gamma passing rates in VMAT prostate external radiation therapy plans.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- A Novel Medical Conformity Index (MCI) Integrating Spatial Distance and Target Volume Weighting for Enhanced Radiotherapy Plan Evaluation.Technology in cancer research & treatment · 2026
- Machine learning prediction of electronic-portal-imaging-device-based in vivo dosimetry results using plan complexity metrics, machine output and gamma passing rate: a feasibility study.Physical and engineering sciences in medicine · 2026