A comparative analysis of pupillometry measurements obtained from different devices featuring static and dynamic capabilities.
Burcu Kazanci, Büşra Dilara Yildirim Erdal, Erdem Topsakal
Department of Ophthalmology, Ankara Etlik City Hospital, Ankara, Turkey. drburcus@hotmail.com. · Department of Ophthalmology, Ankara Etlik City Hospital, Ankara, Turkey.
Key points
From the paper's conclusion- A total of 68 healthy eyes (34 female, 34 male; mean age 27.65 ± 3.84 years) were included. Comparing Metrovision and MS-39, pupil diameters differed significantly under all lighting conditions (all p < 0.001), with good ICC values (photopic: 0.840, mesopic: 0.827, scotopic: 0.786). Bland-Altman analysis showed acceptable agreement only under mesopic conditions (bias: 0.18 mm; p = 0.072), whereas photopic and scotopic measurements were not interchangeable (p < 0.001 and p = 0.005, respectively). Comparing Metrovision and Pentacam AXL Wave, photopic measurements were not interchangeable (bias: - 0.65 mm; p < 0.001), while mesopic measurements showed acceptable agreement (p = 0.053). MS-39 and Pentacam AXL Wave demonstrated interchangeable measurements across mesopic and photopic measurements (mesopic p = 0.933, photopic p = 0.301).
Abstract
PURPOSE: To compare mesopic, scotopic and photopic pupil diameters measured with 2 static and 1 dynamic pupillometry device before refractive surgery. METHODS: We analyzed preoperative measurements from a total of 68 eyes (68 patients) scheduled for corneal refractive surgery. Scanning procedures were performed in the same session, by the same examiner, under standardized conditions. Measurements obtained with the MS-39 Placido Combined Anterior Segment OCT and Pentacam AXL Wave, which are monocular closed-field devices, were compared with those obtained using Metrovision, a binocular open-field device with both dynamic and static pupillometry capabilities. Results were presented using Bland-Altman plots to assess inter-device agreement. RESULTS: A total of 68 healthy eyes (34 female, 34 male; mean age 27.65 ± 3.84 years) were included. Comparing Metrovision and MS-39, pupil diameters differed significantly under all lighting conditions (all p < 0.001), with good ICC values (photopic: 0.840, mesopic: 0.827, scotopic: 0.786). Bland-Altman analysis showed acceptable agreement only under mesopic conditions (bias: 0.18 mm; p = 0.072), whereas photopic and scotopic measurements were not interchangeable (p < 0.001 and p = 0.005, respectively). Comparing Metrovision and Pentacam AXL Wave, photopic measurements were not interchangeable (bias: - 0.65 mm; p < 0.001), while mesopic measurements showed acceptable agreement (p = 0.053). MS-39 and Pentacam AXL Wave demonstrated interchangeable measurements across mesopic and photopic measurements (mesopic p = 0.933, photopic p = 0.301). DISCUSSION: In conclusion, monocular static devices demonstrated relatively better agreement with each other under photopic and mesopic conditions. However, agreement with Metrovision, a binocular device capable of dynamic pupillometry and more representative of physiological pupillary behavior, was condition-dependent and more limited, although compatibility was observed under mesopic conditions. These findings suggest that pupillometry devices should not be considered fully interchangeable across all illumination conditions. While static monocular devices may provide comparable measurements in selected settings, binocular dynamic pupillometry should be considered particularly in patients for whom low-light or scotopic visual performance is clinically important for refractive surgery planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pupil size measurements with two multifunctional devices based on different technologies and an infrared based pupillometer.Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2026
- Comparison of White-to-White Measurements Between Pentacam and MS-39 and Their Relationship with Visual and Refractive Status.Turkish journal of ophthalmology · 2026
- Inter-device Agreement of Device-Designated Chord Mu and Chord Alpha Measurements in Patients Undergoing Preoperative Cataract Evaluation.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Clinical Interchangeability Boundaries of Pentacam AXL WAVE in Preoperative Biometry for Cataract Surgery: An Agreement Study with IOLMaster 700 and Pentacam.Journal of clinical medicine · 2026
- A comparison between two indices derived from automated pupillometry devices.Journal of clinical monitoring and computing · 2026
- Repeatability, Reproducibility and Inter-Device Agreement of Anterior and Posterior Corneal Parameters Measured with Visionix 130+ and Pentacam HR in Healthy EyesPhotonics · 2026