Endoleak detection with photon-counting detector CT at various radiation doses after endovascular aneurysm repair: a phantom study.
Adrian T Huber, Gergely Szucs-Farkas, Ludovic G Perrin, Badis M'Hamdi, Sofia Bespyatova, Andreas Christe, Zsolt Szucs-Farkas
Department of Diagnostic, Interventional and Pediatric Radiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. · University of Bern, Bern, Switzerland.
Key points
From the paper's conclusion- In summary, diagnostic performance for 6 mm endoleaks remained high with 50%-dose reduction using PCD-CT in simulated medium but not in large patients.
Abstract
After endovascular aneurysm repair, patients require lifelong CT follow-ups. Photon-counting detector CT (PCD-CT) may improve detection of tiny iodinated structures at low radiation dose. The aim of this experimental study was to find the lowest dose which still provides high detectability of small endoleaks using PCD-CT with 55 keV and 50 keV virtual monoenergetic images (VMIs). An abdominal aneurysm phantom mimicking medium and large patients was scanned at 100%, 50%, 25%, and 12.5% radiation dose using PCD-CT. Besides standard 55 keV images, 50 keV VMI were reconstructed at 25% and 12.5% dose. The ninety-six CT series containing 1728 endoleaks of 2-6 mm were read by three independent observers. Areas under the alternative free-response operational characteristic curve (A1) were analysed with the jackknife-method. In the medium phantom, A1 for 6 mm endoleaks was similar at 100% and 50% doses (0.93 vs. 0.86; p = 0.23), but performance at 50% dose decreased in 4-6 mm endoleaks (A1, 0.81 vs. 0.68; p = 0.02). In the large phantom, A1 decreased at 50% dose both for 6 mm (0.72 vs. 0.42; p = 0.003) and for 4-6 mm (0.49 vs. 0.29; p = 0.02) endoleaks. Compared with standard reconstruction, 50 keV VMIs did not change A1 at 25% dose (0.56 vs. 0.52; p = 0.38) and led to worse performance at 12.5% dose (0.30 vs. 0.19; p = 0.01). In summary, diagnostic performance for 6 mm endoleaks remained high with 50%-dose reduction using PCD-CT in simulated medium but not in large patients. 50 keV VMIs at 12.5% and 25% dose did not improve endoleak detection.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dual-Source Photon-Counting CT for Thoracic Arteries: Enhancing Image Quality and Diagnostic Assessability with Low-Energy Virtual Monoenergetic Imaging.Diagnostics (Basel, Switzerland) · 2026
- Improvement of Lung Nodule Volumetric Accuracy with Photon-counting Computed Tomography Over Energy-integrating Computed Tomography in Low-dose ScreeningInvestigative Radiology · 2026
- Reducing contrast media dosage for pulmonary embolism CTPA in PCD-CT: a comparative study of EID-CT and PCD-CT in the era of individualized protocollingEuropean Radiology · 2025
- Pulmonary nodule image quality and morphologic visualization with photon-counting detector CT: a scoping reviewEuropean Radiology Cardiovascular and Thoracic · 2026
- Pulmonary Iodine Map Defects Across Contrast Phases in Routine Photon-Counting Detector CT: An Exploratory Study.Diagnostics (Basel, Switzerland) · 2026
- Diagnostic performance of clinical photon-counting CT in the assessment of peri-implantitis - an ex vivo diagnostic accuracy study.International journal of implant dentistry · 2026