Dual-Source Photon-Counting CT for Thoracic Arteries: Enhancing Image Quality and Diagnostic Assessability with Low-Energy Virtual Monoenergetic Imaging.
Andreea I Nica, Christian Booz, Giuseppe M Bucolo, Leona S Alizadeh, Thomas J Vogl, Katrin Eichler, Vitali Koch, Leon D Gruenewald,
Division of Experimental Imaging, Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590 Frankfurt Am Main, Germany. · Diagnostic and Interventional Radiology Unit, BIOMORF Department, University Hospital "Policlinico G. Martino", Via Consolare Valeria 1, 98100 Messina, Italy.
Key points
From the paper's conclusion- Conclusions: VMI reconstructions at low energy levels (40-55 keV) significantly enhance vascular CTvalue and CNR, offering superior image quality and diagnostic assessability for thoracic arteries compared to standard CT series in photon-counting CTA.
Abstract
Background/Objectives: This study aimed to investigate the impact of low-energy virtual monoenergetic imaging (VMI) on quantitative and qualitative image characteristics, as well as its effect on the diagnostic assessability of thoracic arteries in photon-counting computed tomography angiography (CTA). Methods: We retrospectively evaluated 125 patients who underwent dual-source photon-counting CTA scans of thoracic arteries. We reconstructed standard CT images at 120 kV and VMI series at 15 keV intervals, ranging from 40 to 100 keV, and conducted quantitative and qualitative image analyses. For quantitative analysis, we assessed vascular CT numbers and contrast-to-noise ratio (CNR). For qualitative evaluation, three board-certified radiologists independently rated image quality, vascular contrast, and diagnostic assessability of the thoracic arteries using a five-point scale. Results: Quantitative image analysis revealed that 40 keV VMI reconstructions exhibited the highest mean CT value (HU: 1205 ± 286) and CNR (29.22 ± 9.13), followed by the 55 keV series (HU: 679 ± 161, CNR: 22.54 ± 7.36), significantly improved compared to higher keV levels and the standard 120 kV CT series (p < 0.001). Qualitative analysis showed the highest rating scores for 55 keV reconstructions, significantly higher than those of VMI series at higher energy levels and the standard 120 kV series (p < 0.001). Conclusions: VMI reconstructions at low energy levels (40-55 keV) significantly enhance vascular CTvalue and CNR, offering superior image quality and diagnostic assessability for thoracic arteries compared to standard CT series in photon-counting CTA.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Plaque-Adapted Virtual Monoenergetic Image Selection for Carotid Plaque Visualization Using Photon-Counting CT.Diagnostics (Basel, Switzerland) · 2026
- Optimized dual-energy CT protocol for inferior vena cava CT angiography using 60 keV virtual monoenergetic imaging and 60% ASIR-V.Frontiers in cardiovascular medicine · 2026
- Improvement of Lung Nodule Volumetric Accuracy with Photon-counting Computed Tomography Over Energy-integrating Computed Tomography in Low-dose ScreeningInvestigative Radiology · 2026
- Practical applications of photon-counting CT in oncology: an organ-based pictorial review.Insights into imaging · 2026
- Photon-counting-CT – der neue Standard in der kardiovaskulären Diagnostik?Wiener klinisches Magazin · 2026
- Spectral ultrahigh-resolution photon-counting CT for coronary stent imaging: evaluation in a dynamic phantomEuropean Radiology Experimental · 2025