Pulmonary nodule image quality and morphologic visualization with photon-counting detector CT: a scoping review

Felix Barajas Ordonez, Isaac Eli Martínez Safar, Ana M. Rendón-Garavito, Can Yüksel, Christiane Kuhl, Omar Andrés Pantoja-Burbano, Felipe Aluja-Jaramillo

Review

Key points

From the paper's conclusion
  • Findings Photon-counting detector CT showed superior subjective image quality, additional morphologic features in 34%–73% of subsolid nodules, and 12%–18% lower doses versus energy-integrating detector CT.

Abstract

Objective To systematically map the available evidence on pulmonary nodule visualization quality using photon-counting CT (PCD-CT). Materials and methods This scoping review followed JBI methodology and PRISMA-ScR reporting standards. MEDLINE/PubMed, Embase, and Scopus were searched from January 2021 to January 2026. Studies evaluating pulmonary nodule visualization in adults or ex vivo human specimens using PCD-CT were included. Two reviewers independently screened records and extracted data; consistent with scoping review methodology, no formal risk-of-bias assessment was performed. Results Five studies (six records; 318 patients, mean age range 56–69 years) and 20 cadaveric lungs met the inclusion criteria. All studies used the same PCD-CT platform. Image quality: same-day comparisons showed superior subjective image quality with PCD-CT at 12%–18% lower radiation doses ( p < 0.001). Morphologic feature visualization: PCD-CT detected additional features in 34%–73% of subsolid nodules; ex vivo histological correlation demonstrated higher morphological concordance at 1024-matrix resolution. Nodule detection: ultra-low-dose PCD-CT maintained detection sensitivity of 0.94. Conclusion PCD-CT shows promise for improving pulmonary nodule visualization at reduced radiation doses. Evidence remains limited to single-center studies without clinical outcome validation, precluding implementation recommendations. Key Points Question Whether photon-counting detector CT improves image quality and morphologic characterization of pulmonary nodules compared with energy-integrating detector CT remains insufficiently mapped. Findings Photon-counting detector CT showed superior subjective image quality, additional morphologic features in 34%–73% of subsolid nodules, and 12%–18% lower doses versus energy-integrating detector CT. Relevance statement In single-center studies, photon-counting detector CT improved morphologic characterization of pulmonary nodules and reduced radiation exposure; clinical outcome validation is required before broad implementation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Pulmonary nodule image quality and morphologic visualization with photon-counting detector CT: a scoping review — Science Skim