Pediatric radiology

Evaluation of a novel flexible, lightweight blanket-like torso coil in pediatric MRI at 1.5-T - initial experience.

Jonathan R Dillman, Jean A Tkach, R Paul Guillerman, Murat Kocaoglu, Yinan Li, Michael P Nasser, Adam F Prasanphanich, Amol Pednekar,

Department of Radiology, Cincinnati Children's Hospital Medical Center, 3333 Burnet Avenue, Cincinnati, OH, 45229, United States. jonathan.dillman@cchmc.org. · Department of Radiology, Cincinnati Children's Hospital Medical Center, 3333 Burnet Avenue, Cincinnati, OH, 45229, United States.

Observational · n=20

Key points

From the paper's conclusion
  • The flexible blanket-like torso coil provided image quality similar to a conventional coil and was generally preferred by patients and MRI technologists.

Abstract

BACKGROUND: Conventional phased-array torso coils used for abdominal MRI are relatively heavy and inflexible, potentially affecting patient comfort and technologist workflow. OBJECTIVE: To compare image quality, patient experience, and technologist experience using a recently United States Food and Drug Administration-cleared flexible, lightweight, blanket-like torso coil versus a conventional torso coil for pediatric abdominal MRI. MATERIALS AND METHODS: In this prospective institutional review board-approved study, 20 pediatric patients (7 years-old to 18 years-old) undergoing routine clinical outpatient abdominal MRI were imaged on the same 1.5-T system using both a conventional torso coil and a flexible blanket-like torso coil (Smart Fit TorsoCardiac; Philips Healthcare). After standard clinical imaging, selected sequences were repeated following coil exchange. Five pediatric radiologists independently evaluated paired image sets for subjective signal-to-noise ratio (SNR) and coil-related artifacts. Objective image quality was assessed using organ- and tissue-based SNR estimates and contrast-to-noise ratio (CNR) measurements. Patients and MRI technologists completed questionnaires regarding comfort, usability, and examination experience. RESULTS: All 20 patients (median age, 15 years; 55% female) underwent imaging with both coils. Radiologists rated the novel coil as similar to the conventional coil for subjective SNR across all three sequences under study; coil-related artifacts were uncommon (2% of assessments for all sequences evaluated). Objective analysis showed no significant image-based SNR differences for most sequences, although the flexible coil demonstrated higher liver SNR on axial diffusion-weighted imaging (adjusted P=0.01). CNR did not differ between coils. Patients reported greater comfort and less anxiety with the flexible coil; 55% preferred it for future MRI examinations versus 20% for the conventional coil. Technologists also favored the flexible coil, with 70% expressing an overall preference. CONCLUSION: The flexible blanket-like torso coil provided image quality similar to a conventional coil and was generally preferred by patients and MRI technologists.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers