Improved SNR and Image Quality of Pediatric MRI with a Shape Conforming High-Density Posterior Cushion Coil.
Terrence Jao, Praveen Jayapal, Jana Vincent, John Pauly, Ceara Stack, Victor Taracila, Fraser Robb, Thomas Grafendorfer,
Department of Radiology, Stanford University, 300 Pasteur Ave., Stanford, CA 94305, USA. · General Electric Healthcare, Aurora, OH 44202, USA.
Key points
From the paper's conclusion- Our goal is to improve pediatric MRI by constructing a prototype 30-channel high-density posterior cushion coil (HDPCC) incorporated into a padded cushion.
Abstract
A high-density posterior cushion coil (HDPCC) can improve the SNR in pediatric MRI by conforming to the body and minimizing distance to the imaged volume when compared to a conventional rigid built-in table coil array. Posterior coil arrays built into MRI tables are mainstays but provide suboptimal performance in pediatric imaging due to their distance from the imaged volume and lower coil density. Our goal is to improve pediatric MRI by constructing a prototype 30-channel high-density posterior cushion coil (HDPCC) incorporated into a padded cushion. Twenty-five pediatric patients were scanned with the HDPCC, and twenty-five pediatric controls served as the comparison group using standard coils. Two radiologists graded image quality on a 5-point scale. SNR was measured in the right paraspinal musculature and the liver. All images obtained with the HDPCC were diagnostic (mean score 4.13/5). SNR in the superficial tissues of the paraspinal musculature was 34% higher with the HDPCC than with standard coils, while SNR in the deeper tissues of the liver was statistically similar between the two coils. HDPCC yields improved the signal-to-noise ratio in phantoms and pediatric study participants by conforming to the body.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of a novel flexible, lightweight blanket-like torso coil in pediatric MRI at 1.5-T - initial experience.Pediatric radiology · 2026
- Pediatric thoracic magnetic resonance imaging: current evidence, technical advances, and clinical integration.Diagnostic and interventional radiology (Ankara, Turkey) · 2026
- Augmented 3T imaging performance of the Temporomandibular joint using wireless resonators.Journal of magnetic resonance (San Diego, Calif. : 1997) · 2026
- Phantom imaging performance and initial in vivo feasibility of an anatomy-conforming dentomaxillofacial prototype coil for dental MRIBMC Medical Imaging · 2026
- Integrating Low-loss High-permittivity Ceramic Helmet with RF Head Array Coil to Largely Enhance Coil Sensitivity and SNR for Human Brain MRI at 7T.IEEE transactions on bio-medical engineering · 2026
- Highly Superficial Flexible Coil Array for 1.5T MRI: Diagnostic Quality Evaluation Compared to Conventional Technology.IEEE transactions on bio-medical engineering · 2026