5.0 T versus 1.5 T and 3.0 T upper abdominal MRI: a prospective two‑cohort study of image quality and diagnostic confidence.
Liuhong Zhu, Danlan Lian, Funan Wang, Anjie Xie, Jiaqi Wang, Hao Liu, Jianjun Zhou
Department of Radiology, Zhongshan Hospital (Xiamen), Fudan University, Xiamen, China. · Collaborative Innovation Department, United Imaging Healthcare, Shanghai, China.
Key points
From the paper's conclusion- 5.0 T upper abdominal MRI, with a primary focus on liver imaging, is clinically feasible, offering superior anatomical detail, SNR, overall image quality, and diagnostic confidence over 1.5 T and 3.0 T, though cardiac motion artifacts, liver signal inhomogeneity on T2WI, and geometric distortion on DWI require further optimization.
Abstract
OBJECTIVES: To compare 5.0 T magnetic resonance imaging (MRI) image quality versus 1.5 T and 3.0 T in patients and healthy volunteers, so as to validate its clinical feasibility. MATERIALS AND METHODS: This prospective study included a patient cohort (348 consecutive patients undergoing upper abdominal MRI) and a healthy volunteer cohort (31 volunteers scanned on all three scanners). Two radiologists independently assessed image quality subjectively (5-point Likert scale: anatomical detail, artifacts, overall image quality, diagnostic confidence) and measured signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Subjective and objective metrics were compared across field strengths using overall and pairwise comparisons. RESULTS: Objective metrics (left/right/mean liver SNR, muscle/spleen SNR, liver-spleen CNR) showed a significant gradient of 5.0 T > 3.0 T > 1.5 T in both cohorts (all p < 0.001). Subjectively, compared with 1.5 T and 3.0 T, 5.0 T received the lowest scores for cardiac motion artifact and liver signal uniformity on T2-weighted imaging (T2WI) and showed greater distortion and ghosting artifacts on diffusion-weighted imaging (DWI) (all p < 0.05). Nevertheless, 5.0 T was significantly superior in anatomical detail, SNR, and overall image quality (all p < 0.05). In patients, 5.0 T also achieved the highest diagnostic confidence (p < 0.05). Results from healthy volunteers were largely consistent with those from the patient cohort. CONCLUSION: 5.0 T upper abdominal MRI, with a primary focus on liver imaging, is clinically feasible, offering superior anatomical detail, SNR, overall image quality, and diagnostic confidence over 1.5 T and 3.0 T, though cardiac motion artifacts, liver signal inhomogeneity on T2WI, and geometric distortion on DWI require further optimization. KEY POINTS: Question Is 5.0-T MRI superior to 1.5 T and 3.0 T for liver imaging, and does its diagnostic benefit differ between diseased and non-diseased livers? Findings 5.0 T provided significantly higher SNR, image quality, and diagnostic confidence than 1.5 T and 3.0 T, despite increased cardiac motion and liver signal inhomogeneity. Clinical relevance 5.0-T MRI significantly improves liver image quality and diagnostic confidence over 3.0 T and 1.5 T, supporting its clinical utility despite increased artifacts that require further optimization.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Non-contrast renal MR angiography: a within-subject comparison of imaging performance at 1.5 T, 3 T, and 5 T.BMC medical imaging · 2026
- Diagnostic image quality of shoulder magnetic resonance imaging at 0.55T in healthy subjects and patients with rotator cuff injury: Prospective comparison with 1.5TJournal of Applied Clinical Medical Physics · 2026
- Highly Superficial Flexible Coil Array for 1.5T MRI: Diagnostic Quality Evaluation Compared to Conventional Technology.IEEE transactions on bio-medical engineering · 2026
- Effect of Visualized Respiratory Training on Patient Cooperation and Image Quality in Gadoxetic Acid Disodium‐Enhanced Liver MRI : A Randomized Controlled TrialJournal of Magnetic Resonance Imaging · 2026
- 5T Non-Contrast One-Stop MRA for Thoracic Aorta and Coronary Arteries: A Prospective Feasibility Study in Healthy Volunteers.Journal of cardiovascular development and disease · 2026
- Diagnostic Performance of a Metasurface Coil Compared With a Commercial 12-Channel Coil for Knee MRI at 1.5 T.Journal of magnetic resonance imaging : JMRI · 2026