Bridging the Gap Between Chemiluminescence Immunoassay and Liquid Chromatography-Tandem Mass Spectrometry for Infliximab Monitoring: Method Validation and Clinical Comparability Study.
Ying Xia, Ming-Mei Zhu, Meng-Qing Xiao, Chen-Chun Zhong, Yu-Jun Zhou, Yi-Hao Xue, Shi-Wei Zheng, Li Li,
Department of Pharmacy, Children's Hospital of Nanjing Medical University, Nanjing, China. · Shanghai AB Sciex Co., Ltd., Shanghai, China.
Key points
From the paper's conclusion- This study aimed to validate an in-house chemiluminescence immunoassay (CLIA) for IFX quantification and evaluate its agreement with liquid chromatography-tandem mass spectrometry (LC-MS/MS).
Abstract
Infliximab (IFX) is a cornerstone biologic for pediatric inflammatory bowel disease, where therapeutic drug monitoring (TDM) is essential but complicated by inter-platform variability. Systematic comparison of automated immunoassays with reference methods is therefore warranted. This study aimed to validate an in-house chemiluminescence immunoassay (CLIA) for IFX quantification and evaluate its agreement with liquid chromatography-tandem mass spectrometry (LC-MS/MS). A CLIA kit was validated by assessing the limit of blank, specificity, linearity, accuracy, precision, matrix equivalence, and stability. IFX concentrations in 52 pediatric plasma samples were measured by CLIA and a reference LC-MS/MS method. Comparability was evaluated using linear regression, Passing-Bablok regression, and Bland-Altman analysis. CLIA showed excellent performance (lower limit of quantification: 0.640 µg/mL; specificity <10.0% interference; reportable range: 0.640-462 µg/mL). LC-MS/MS yielded higher concentrations than the CLIA (median: 6.25 vs. 2.87 µg/mL). Strong correlation was observed (r = 0.9101) with no significant deviation from linearity (Passing-Bablok, p > 0.05). However, Bland-Altman revealed substantial mean relative bias (67.9%) and poor categorical agreement (52% concordance) within the therapeutic range (3-7 µg/mL), indicating non-interchangeability. We validated a rapid, automated CLIA suitable for high-throughput TDM. Despite strong correlation, significant quantitative differences preclude direct result substitution, underscoring the need for method-specific therapeutic thresholds.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-408 Evaluation of the infliximab and adalimumab therapeutic drug monitoring tests from ProciseDx in a pediatric settingClinical Chemistry · 2026
- Simultaneous Quantification of Vedolizumab, Ustekinumab, and Infliximab in Human Serum by LC-MS/MS: Application to Therapeutic Drug Monitoring in Inflammatory Bowel Disease.Drug design, development and therapy · 2026
- Comparison of LC-MS/MS and CLIA Methods for Vitamin D Measurement in an Unselected Outpatient Cohort: Statistical Evaluation and Impact on Clinical Classification.Diagnostics (Basel, Switzerland) · 2026
- Connecting the Dots Between Pharmacokinetics, Inflammatory Biomarkers, and Mucosal Healing in Patients on Infliximab with Inflammatory Bowel Disease.Pharmaceutics · 2026
- Beyond correlation: Intraclass correlation and bland-altman analysis reveal poor absolute agreement across multiplex cytokine immunoassay platforms.Cytokine · 2026
- B-343 LC MS/MS Measurement of Infliximab, Adalimumab, Ustekinumab, Vedolizumab and Anti Drug Antibodies Using mAbXmise Kits on Xevo TQ Absolute XR MS for Clinical ResearchClinical Chemistry · 2026