Comparison of LC-MS/MS and CLIA Methods for Vitamin D Measurement in an Unselected Outpatient Cohort: Statistical Evaluation and Impact on Clinical Classification.
Carlo Corbetta, Luigi Corsaro, Eugenio Caradonna, Chiara Rusconi, Ivana De Rosa, Stefano Diani, Veniero Gambaro, Ugo de Grazia,
Laboratory Research Unit, CDI-Centro Diagnostico Italiano S.p.A., Via Saint Bon 20, 20147 Milan, Italy. · Medical Genetic Laboratory, CDI-Centro Diagnostico Italiano S.p.A., Via Saint Bon 20, 20147 Milan, Italy.
Key points
From the paper's conclusion- In our study, despite preserved rank ordering, a non-linear concentration-dependent bias and different limits of agreement caused systematic differences in the classification of vitamin D deficiency. Adoption of automated LC-MS/MS can substantially reclassify outpatients toward sufficiency, with implications for supplementation and laboratory harmonisation.
Abstract
BACKGROUND: This study aimed to compare a chemiluminescence immunoassay (CLIA; Siemens Atellica IM 1600) with an automated liquid chromatography-tandem mass spectrometry (LC-MS/MS) analyser (Thermo Scientific Cascadion SM) for serum 25-hydroxyvitamin D [25(OH)D] measurement in unselected outpatients, and its impact on clinical classification. METHODS: 25(OH)D was measured on 1064 paired serum samples (828 women, 236 men; age 3-111 years); both platforms participated in the Vitamin D Standardisation Program (VDSP). Agreement was assessed by the Wilcoxon signed-rank test, Passing-Bablok regression, Bland-Altman analysis on log-transformed values, Breusch-Pagan testing for heteroscedasticity, consistency-based intraclass correlation coefficient (ICC), Lin's concordance correlation coefficient, and Cohen's kappa with McNemar's test on six- and two-category clinical classifications. RESULTS: Methods differed significantly (p<2.2×10-16). Passing-Bablok regression showed proportional bias (slope 1.33; 95% confidence interval [CI], 1.28-1.38; LC-MS/MS higher). Consistency ICC was 0.928 (95% CI, 0.919-0.936). Log-Bland-Altman bias was 35.5% (limits of agreement, -18.3% to +124.5%; heteroscedasticity, p=5.18×10-11). Cohen's kappa was 0.32 unweighted and 0.72 weighted on six clinical categories, and 0.56 on a two-class scheme (McNemar p<2.2×10-16). In discordant pairs, CLIA classified deficiency 118-fold more often than LC-MS/MS. CONCLUSIONS: In our study, despite preserved rank ordering, a non-linear concentration-dependent bias and different limits of agreement caused systematic differences in the classification of vitamin D deficiency. Adoption of automated LC-MS/MS can substantially reclassify outpatients toward sufficiency, with implications for supplementation and laboratory harmonisation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of a new non-competitive chemiluminescent immunoassay for 25-OH vitamin D measurement.Diagnosis (Berlin, Germany) · 2026
- Comprehensive Comparison of Three Automated 25-Hydroxyvitamin D Immunoassays with a VDSCP Traceable LC-MS/MS Method Across Clinically Relevant Decision Thresholds.Nutrients · 2026
- Is a low-cost latex immunoturbidimetric assay suitable for routine 25(OH)D screening? Analytical comparison with chemiluminescent immunoassay and LC-MS/MS.Scandinavian journal of clinical and laboratory investigation · 2026
- B-029 Analytical Performance Evaluation of the Double-Antibody Sandwich Mindray Total-Vitamin D Assay and Comparison Versus Liquid Chromatography Tandem Mass Spectrometry and Commercial ImmunoassaysClinical Chemistry · 2026
- Analytical performance of a single epitope meta-sandwich total 25-hydroxyvitamin D immunoassay and establishment of reference intervals in a Chinese population.Annals of clinical biochemistry · 2026
- B-365 Breaking Throughput Barriers: Automated Vitamin D Testing on the Roche Cobas i601 LC-MS/MS in a Tertiary Hospital LaboratoryClinical Chemistry · 2026