Beyond correlation: Intraclass correlation and bland-altman analysis reveal poor absolute agreement across multiplex cytokine immunoassay platforms.
Olfa Khalfallah, Cynthia Le Mouroux, Narjis Kraimi, Claire Sendra, Nathalie Saint-Laurent, Maryline Guidetti, Laurence Pellegrina, Fanchon Herman,
Institut de Pharmacologie Moléculaire et Cellulaire (IPMC), CNRS, INSERM, Université Côte d'Azur, 660 Route Des Lucioles, 06560 Valbonne, France. Electronic address: Khalfallah@ipmc.cnrs.fr. · Institut de Pharmacologie Moléculaire et Cellulaire (IPMC), CNRS, INSERM, Université Côte d'Azur, 660 Route Des Lucioles, 06560 Valbonne, France. Electronic address: cynthia-lm@hotmail.com.
Key points
From the paper's conclusion- These findings show that OLINK, LUMINEX, and V-PLEX measurements are not interchangeable at the level of absolute concentrations and underscore key analytical limitations that currently hinder the broad adoption of cytokine immunoassays, particularly multiplex panels, as regulator-approved tests.
Abstract
Cross-platform comparisons of serum cytokine immunoassays have often focused on rank correlation, but the extent of absolute agreement and systematic bias between platforms remains insufficiently characterized. Few independent head-to-head evaluations have compared multiple immunoassay platforms on the same clinical samples using overlapping cytokines. Here, serum samples from 48 adult participants were analyzed in parallel for seven cytokines (IL-6, IL-7, IL-8, IL-10, IL-15, TNF-α, IFN-γ), four chemokines (CCL2, CCL3, CCL4, CXCL10), and one growth factor (VEGF-A) using three multiplex platforms: OLINK proximity extension assay-based, LUMINEX bead-based, and V-PLEX electrochemiluminescence-based immunoassays. LLOQs, absolute concentrations, rank correlations, ICCs, and Bland-Altman biases were compared between platforms. Absolute serum cytokine concentrations differed by up to two orders of magnitude. Correlations between OLINK and V-PLEX were generally strong, whereas correlations involving LUMINEX were often weak or non-significant. ICCs were poor (<0.5) for most analytes and platform pairs, with IL-8 as a notable exception. Bland-Altman analyses on log-transformed concentrations revealed large, analyte-specific multiplicative biases and wide ratio-scale limits of agreement, indicating poor absolute agreement. These findings show that OLINK, LUMINEX, and V-PLEX measurements are not interchangeable at the level of absolute concentrations and underscore key analytical limitations that currently hinder the broad adoption of cytokine immunoassays, particularly multiplex panels, as regulator-approved tests.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Bridging the Gap Between Chemiluminescence Immunoassay and Liquid Chromatography-Tandem Mass Spectrometry for Infliximab Monitoring: Method Validation and Clinical Comparability Study.Journal of separation science · 2026
- Benchmarking batch-effect correction in clinical bead-based multiplex immunoassays: a framework and web tool for cytokine and soluble immune checkpoint data.Frontiers in immunology · 2026
- PFLISA: A New Ultrasensitive, Multiplex Immunoassay Platform Enables Cytokine Network Profiling in SepsisbioRxiv · 2026
- Bridging variability: toward standardized and harmonized multiplex bead-based flow cytometric cytokine assaysClinical Chemistry and Laboratory Medicine (CCLM) · 2026
- Cytokine profiling of molecular endotypes of knee osteoarthritis: insights from the IMI-APPROACH cohortArthritis Research & Therapy · 2026
- A Comprehensive Mathematical Model of Avidity in Cytokine SignalingbioRxiv · 2026