The suitability of the EPIbeam system for dosimetric verification of radiotherapy treatment plans
Aleksandra Żaba, Andrzej Chmiel, Katarzyna Matusiak
Key points
From the paper's conclusion- Conclusions Despite certain limitations, EPIbeam proved to be a reliable tool for dosimetric verification of treatment plans implemented using the dynamic VMAT technique.
Abstract
Introduction The aim of this study was to validate the DOSIsoft’s EPIbeam portal dosimetry system for controlling dose distribution in radiotherapy delivered by dynamic techniques. Material and methods Treatment plans were verified using multidetector matrices and an EPID dosimetry panel. Additionally, functional tests and simulations of potential errors that may occur during irradiation were performed using the EPID panel. The comparison of dose distributions planned in the treatment planning system with the distributions obtained as a result of the application of two verification methods was performed in dedicated software using the gamma coefficient analysis method. A statistical analysis of the obtained results was performed to determine the most advantageous EPIbeam system settings and the analysis criteria to be applied. The EPIbeam system was then compared with previously used system. Results Statistical analysis indicated that one of the three accelerator model libraries created in the EPIbeam system (library 15, DLS = 0.4 mm) should be excluded from clinical use. Significant differences were observed for this library when applying the 2%, 2 mm gamma criterion. When applying the 2%, 2 mm criterion with the accelerator model set in library 8 (DLS = 0.0 mm), the results show no differences compared to multidetector matrices. This indicated that verification performed for these settings in the EPIbeam system yielded results similar to verification performed using matrices. Conclusions Despite certain limitations, EPIbeam proved to be a reliable tool for dosimetric verification of treatment plans implemented using the dynamic VMAT technique.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Sensitivity in error detection of an electronic portal imaging device for patient-specific QA in head and neck VMAT plans.Radiological physics and technology · 2026
- Evaluation of a multifunctional phantom for geometric, image quality, and dosimetric constancy of megavoltage electronic portal imaging devicesJournal of Applied Clinical Medical Physics · 2026
- Retrospective analysis of machine service logs and EPID-based linac QA to investigate the effects of machine maintenance and calibration events on beam characteristics: A multi-institutional study.Journal of applied clinical medical physics · 2026
- Assessing Gamma Index Sensitivity to Selected Linear Accelerator Mechanical Errors Using Various Dosimetric Systems.Journal of clinical medicine · 2026
- Configuration of a Monte Carlo-Based Secondary Plan Verification Tool: An Analytical Approach to User-Specific Parameter Tuning.Biomedical physics & engineering express · 2026
- Independent Three-dimensional Dose Calculation from Log Files for Breast VMAT: A Measurement-free Quality Assurance Approach.Journal of medical physics · 2026