A practical framework for routine verification of Spot Position Monitor coordinate consistency in scanned carbon-ion radiotherapy.
Toshiro Tsubouchi, Noriaki Hamatani, Akira Komada, Masaaki Takashina
Department of Radiation Oncology, Piedmont Mountainside Hospital, Jasper, Georgia, USA. · Department of Medical Physics, Osaka Heavy Ion Therapy Center, Osaka, Japan.
Key points
From the paper's conclusion- The proposed framework provides a practical approach for routine verification of SPM coordinate consistency and for establishing facility-specific baselines through longitudinal monitoring. The RMS residual should not be interpreted as the intrinsic positional error of the SPM alone, but to provide a framework for establishing a facility-specific baseline for monitoring the agreement between SPM-reported and independently measured spot coordinates under routine QA conditions.
Abstract
BACKGROUND: Log file-based patient specific quality assurance (PSQA) in scanned particle therapy relies on the consistency of spot position information recorded by the spot position monitor (SPM), but practical methods for routine independent verification of these coordinates remain limited. PURPOSE: This study describes a practical framework for routine verification of SPM coordinate consistency using independent scintillator-based measurements in raster-scanning carbon-ion radiotherapy. METHODS: Reference spot positions were measured at the isocenter plane using a scintillator-based detector. Corresponding SPM-recorded spot coordinates were extracted from delivery log files. The two coordinate sets were aligned using a similarity transformation accounting for translation, rotation, and isotropic scaling. Post-transformation root-mean-square (RMS) residuals were used as a practical consistency metric reflecting the agreement between SPM-reported and independently measured spot positions under routine QA conditions. RESULTS: After similarity transformation, the median RMS residual across all investigated accelerator energies and irradiation ports was 1.45 mm, with an interquartile range of 1.36-1.63 mm. No statistically significant difference was detected across accelerator energies, whereas a significant difference was observed among irradiation ports. CONCLUSIONS: The proposed framework provides a practical approach for routine verification of SPM coordinate consistency and for establishing facility-specific baselines through longitudinal monitoring. The RMS residual should not be interpreted as the intrinsic positional error of the SPM alone, but to provide a framework for establishing a facility-specific baseline for monitoring the agreement between SPM-reported and independently measured spot coordinates under routine QA conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Site-specific performance and complexity-dependent risk assessment of patient-specific quality assurance using a 2D array detector in CyberKnife radiosurgery.Radiological physics and technology · 2026
- Feasibility study of secondary-electron-bremsstrahlung imaging with a knife-edge slit camera for range verification in carbon-ion therapyJournal of Instrumentation · 2026
- Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel DetectorsarXiv · 2026
- A patient-specific Winston-Lutz system for multi-target, single isocentre radiotherapy.Biomedical physics & engineering express · 2026
- Measuring Small Collimated Proton Fields with a Volumetric Scintillation DetectorJournal of Physics: Conference Series · 2026
- Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR BrachytherapyarXiv · 2026