Muse 2 forehead photoplethysmography agrees with Polar H10 for heart rate but overestimates ultra-short RMSSD
I. P. A. Griadhi, I. A. A. Suadnyana, N. Wahyuni
Key points
From the paper's conclusion- Muse 2 PPG can supply mean heart rate and, with a stated bias, SDNN; its ultra-short RMSSD is not interchangeable with Polar RMSSD and should be reported as pulse-rate variability.
Abstract
The Muse 2 EEG headset also records forehead photoplethysmography (PPG). We tested whether its pulse intervals can replace Polar H10 RR intervals for ultra-short heart-rate variability in 28 seated recordings (60 s eyes open, then 60 s eyes closed), using fixed processing, no exclusions and eleven sensitivity analyses. Mean heart rate agreed almost perfectly (ICC [≥] 0.999; bias -0.22 and 0.00 beats min-1). SDNN agreement was good to excellent (ICC 0.86 and 0.91; bias +8.6 and +8.2 ms). RMSSD agreement was moderate (ICC 0.60): Muse exceeded Polar in at least 27 of 28 sessions (bias +19.2 and +18.5 ms; about 1.9-fold), and the ratio bias fell as RMSSD rose. Interval differences (SD 23 ms) implied about 16 ms of independent timing error per pulse, which reproduced the RMSSD bias; without ECG, sensor jitter could not be separated from variation in pulse transit time. Tracking of the eyes-closed change in ln RMSSD was not demonstrated ({rho} = 0.35, 95% CI -0.07 to 0.67; detectable {rho} [≥] 0.51). Muse 2 PPG can supply mean heart rate and, with a stated bias, SDNN; its ultra-short RMSSD is not interchangeable with Polar RMSSD and should be reported as pulse-rate variability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy of Photoplethysmography-Derived Pulse Rate Variability Compared with Electrocardiography-Derived Heart Rate Variability: A Systematic Review and Meta-Analysis.Sensors (Basel, Switzerland) · 2026
- Method comparison of a ring-based PPG device and a chest-strap monitor for heart rate and heart rate variability in shooting athletes.BMC sports science, medicine & rehabilitation · 2026
- Pulse vs. R-R Interval: Validity and Reliability of the Garmin Elevate v4 Sensor for Ultra-Short HR and HRV Monitoring in Healthy Adults.Sensors (Basel, Switzerland) · 2026
- Contactless camera-based estimation of heart rate and variability from short facial video using remote photoplethysmography, remote ballistocardiography, and quality-aware selectionEuropean Heart Journal · 2026
- Estimating heart rate variability using facial video photoplethysmography: a pilot validation studyScientific Reports · 2026
- Cross-polarised temple photoplethysmography for motion-robust and equitable cardiovascular sensing in virtual reality.Biomedical optics express · 2026