Attributes Affecting Rigid Needle Shield Pull-Off Force and Cap Removal Force in Autoinjector Systems.
Emran Lallow, Zhaoxi Zheng, Michasel Galluppi, Pei-Yuan Hsiao, Yousef Gaber, Nicole Zajac, Kevin McElhenney, Peng Li,
Device Development and Technology, Merck & Co., Inc., Rahway, NJ 07065, United States emran.lallow@merck.com guangli.hu@merck.com. · Device Development and Technology, Merck & Co., Inc., Rahway, NJ 07065, United StatesÂ.
Key points
From the paper's conclusion- Together, these results show that device-level CRF cannot be interpreted from syringe-level POF alone without accounting for test speed, assembly-induced interface conditioning, and material or aging effects.
Abstract
Autoinjector cap removal force (CRF) must be controlled to support reliable cap removal and ease of use. To understand this force, syringe-level rigid needle shield (RNS) pull-off force (POF) must be considered together with the manufacturing and test attributes that can change the RNS interface before or during cap removal. This study evaluated how pull-off test speed, assembly-induced mechanical disturbance, applied axial preload, and elastomer siliconization affect POF and CRF. POF increased linearly with test speed at low-to-intermediate rates before reaching a plateau, showing that measured POF depends on the test rate. Autoinjector assembly reduced device-level CRF relative to syringe-level POF, while applied axial preload similarly reduced subsequent syringe-level POF, indicating that interface conditioning during assembly can alter the relationship between syringe-level POF and device-level CRF. Additional silicone oil modestly decreased POF, whereas one-month room-temperature aging after preload and 60°C/18h heat exposure after recapping each produced only a small additional increase under the evaluated conditions. Together, these results show that device-level CRF cannot be interpreted from syringe-level POF alone without accounting for test speed, assembly-induced interface conditioning, and material or aging effects.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantifying User-Applied Forces on Prefilled Syringe Luer Cone: A Human Factors Ergonomics Study.PDA journal of pharmaceutical science and technology · 2026
- Characterization of an electromagnetic impact-driven needle-free injector with enhanced pulsed control in a low-jet-power regime.Biomedical physics & engineering express · 2026
- A Removable Mechanical Interface for Non-Invasive Strain Measurement in Paddle Shafts: Iterative Design and Experimental Characterization.Sensors (Basel, Switzerland) · 2026
- Comparing a Fixed and Freely Rotating Capsule in a Dry Powder Inhaler Under Realistic Breathing Conditions.Pharmaceutical research · 2026
- Force-reduction of alginate-coated cochlear implant electrode in porcine specimensCurrent Directions in Biomedical Engineering · 2026