Comparing a Fixed and Freely Rotating Capsule in a Dry Powder Inhaler Under Realistic Breathing Conditions.
Karan Shrivastava, Ajit Kumar, Somashekhar S Hiremath, Agisilaos Kourmatzis, Gajendra Singh
Department of Mechanical Engineering, Indian Institute of Technology Madras, Chennai, Tamil Nadu, 600036, India. · School of Mechanical and Materials Engineering, IIT Mandi, Mandi, HP, 175075, India.
Key points
From the paper's conclusion- Fixed-axis rotation significantly outperformed free-axis rotation (~ 6 × higher RPM and lower powder retention). Higher flow rates and faster-accelerating transient profiles increased RPM and collision frequency, enhancing drug release. Fixing the rotational axis of the capsule had the greatest benefit for the slow inhalation profile. The findings offer key insights for the design of future capsule-based DPIs.
Abstract
PURPOSE: To investigate the performance of a capsule-based dry powder inhaler (DPI) under two device configurations - free-axis and fixed-axis capsule rotation - across steady-state and transient inhalation profiles. METHOD: Two new research-based transparent capsule-based DPI were developed-one with a fixed capsule rotational axis and another with a floating capsule configuration. High-speed imaging was used to capture capsule motion, and the effects of flow regimes (laminar, transient, and turbulent), capsule rotation, and breathing profiles on capsule RPM, wall collisions, and powder retention were analyzed. CONCLUSIONS: Fixed-axis rotation significantly outperformed free-axis rotation (~ 6 × higher RPM and lower powder retention). Higher flow rates and faster-accelerating transient profiles increased RPM and collision frequency, enhancing drug release. Fixing the rotational axis of the capsule had the greatest benefit for the slow inhalation profile. The findings offer key insights for the design of future capsule-based DPIs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- CFD-DEM simulation of capsule vibration facilitated powder discharge and dispersion in Breezhaler.International journal of pharmaceutics · 2026
- Revealing Airflow-Particle Dynamics in Dry Powder Inhalers.Small (Weinheim an der Bergstrasse, Germany) · 2026
- Relating in-device particle behaviour to aerosol metrics using high-speed imaging and NGI measurements.International journal of pharmaceutics · 2026
- Advancing Dry Powder Inhalers: A Complete Workflow for Carrier-Based Formulation DevelopmentPharmaceutics · 2026
- Attributes Affecting Rigid Needle Shield Pull-Off Force and Cap Removal Force in Autoinjector Systems.PDA journal of pharmaceutical science and technology · 2026
- Transitioning from Impinging Jets to Single-Jet Impaction: Investigation of an Alternative Nozzle Design for the Respimat® Soft Mist Inhaler.European journal of pharmaceutical sciences : official journal of the European Federation for Pharmaceutical Sciences · 2026