Characterization of an electromagnetic impact-driven needle-free injector with enhanced pulsed control in a low-jet-power regime.
Prachya Mukda, Pichet Ninduangdee, Kulachate Pianthong, Wirapan Seehanam
Phetchaburi Rajabhat University, 38 Moo 8 Department of Mechanical Engineering, Faculty of Engineering and Industrial Technology, Phetchaburi , 76000, Thailand. · Department of Mechanical Engineering, Phetchaburi Rajabhat University, 38 Moo 8 Department of Mechanical Engineering, Faculty of Engineering and Industrial Technology, Phetchaburi , 76000, Thailand.
Key points
From the paper's conclusion- This study elucidates the development and performance characterization of an advanced electromagnetic impact-driven jet injector version 3 (IDM-v3), specifically engineered to achieve controllable jet formation within a low-jet-power regime.
Abstract
This study elucidates the development and performance characterization of an advanced electromagnetic impact-driven jet injector version 3 (IDM-v3), specifically engineered to achieve controllable jet formation within a low-jet-power regime. To overcome the excessive kinetic energy challenges observed in its predecessor (IDM-v2), several critical technical refinements were implemented, including the integration of a high-force solenoid, a calibrated 94 g projectile, and a reduced maximum impact stroke of 6 mm. Experimental results indicate that impact force, the calculated impact peak pressure metric, and exit jet velocity are highly sensitive to the impact configuration. The 0.2 mm nozzle produced the highest impact peak pressure, P_peak, of approximately 206 MPa, calculated from the measured peak target force using P_peak=F_peak/A_n. The corresponding jet power, calculated separately from the measured exit jet velocity, was approximately 2 W. In contrast, the 0.4 mm nozzle produced a higher calculated jet power of approximately 6 W because of its larger nozzle-orifice area. High-speed temporal analysis revealed a multi-pulse response in the area-normalized target-force signal and stable jet morphology within the initial millisecond of actuation. Validation through penetration assays on polyacrylamide gels and fresh ex vivo porcine tissue demonstrated that the 0.2 mm nozzle produced a narrower and more focused penetration channel, whereas the 0.4 mm nozzle produced greater lateral spreading and reached a greater final penetration depth under the tested conditions. These findings characterize the tunability of the IDM-v3 engineering prototype and provide a basis for further experimental development and evaluation. They do not establish biological safety or clinical or veterinary readiness.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Gas-driven needle-free dental injection enables tunable and predictable soft-tissue penetration: high-speed kinematic and micro-computed tomography validation.BMC oral health · 2026
- A Needle-Free Jet Injection System for Safe and Effective Intratympanic Delivery.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Attributes Affecting Rigid Needle Shield Pull-Off Force and Cap Removal Force in Autoinjector Systems.PDA journal of pharmaceutical science and technology · 2026
- Acoustic control of Rayleigh-regime liquid-jet breakup using a through-hole MHz transducerarXiv · 2026
- Dynamic tapered nozzle and multiple waveform modulation enable high-speed piezoelectric jetting without satellite droplets across broad fluid property ranges.Lab on a chip · 2026
- Study on the Jet Evolution Mechanism and Prediction Model of the Dimensionless Limiting Jet Length for Pneumatic Drop-on-Demand Metal Sn DropletFluids · 2026