Mass spectrometry (Tokyo, Japan)

Compensation of Time-of-Flight Drift Using an Internal Reference Ion during Long-Duration LIMAS Depth Profiling.

Ken-Ichi Bajo

Department of Earth and Planetary Sciences, Faculty of Science, Hokkaido University, Sapporo 060-0810, Japan.

Key points

From the paper's conclusion
  • These results demonstrate that the 24Mg2+ peak can be used as an internal reference for TOF drift compensation during long-duration measurements, providing a robust approach for micro- to nanoscale noble gas depth profiling using LIMAS.

Abstract

In this study, the temporal behavior and compensation of time-of-flight (TOF) drift during long-duration depth profiling using Laser Ionization Mass Nanoscope (LIMAS) were investigated. Noble gas depth profiling by LIMAS requires detection of weak isotope signals with nanometer-scale depth resolution. However, such measurements often take several hours to a couple of days, during which TOF peak-position drift and statistical uncertainty may affect analytical precision. In this study, a 4He-implanted olivine substrate was analyzed as a test sample. The TOF spectra were processed by depth-layer stacking, peak extraction, and peak-position analysis. The 24Mg2+ peak was used as a matrix-derived internal reference to monitor and compensate for TOF drift. The peak-position shifts of measured ions showed clear linear correlations with the 24Mg2+ peak shift, indicating that the observed TOF drift is largely systematic. A 24Mg2+-based compensation reduced the peak-position fluctuations of the other ion species to less than 1 ns and improved the mass resolving power. These results demonstrate that the 24Mg2+ peak can be used as an internal reference for TOF drift compensation during long-duration measurements, providing a robust approach for micro- to nanoscale noble gas depth profiling using LIMAS.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers