Cuffless 24-hour ambulatory blood pressure monitoring-guided antihypertensive titration in real-world clinical practice: a prospective non-randomized study.
Jung-Joon Cha, Mi-Na Kim, So Ree Kim, Jae Hyoung Park, Hyung Joon Joo, Cheol Woong Yu
Division of Cardiology, Department of Internal Medicine, Korea University Anam Hospital, Korea University College of Medicine, Seoul, Republic of Korea.
Key points
From the paper's conclusion- In this non-randomized study, a care pathway incorporating cuffless 24-hour ABPM was associated with lower 24-hour ambulatory systolic BP at 6 months compared with usual care, whereas office systolic BP did not differ significantly between arms. Given the non-randomized design, these findings are hypothesis-generating and require confirmation in a randomized trial.
Abstract
BACKGROUND: Twenty-four-hour ambulatory blood pressure monitoring (ABPM) is the reference standard for diagnosing hypertension and assessing treatment response, but routine use remains limited by conventional cuff-based devices. We compared 6-month ambulatory blood pressure (BP) between cuffless 24-hour ABPM-guided titration and usual care in outpatient hypertension management. METHODS: We conducted a single-center, open-label, non-randomized prospective study of adults with uncontrolled hypertension between March and November 2025. Participants were assigned by the treating physician, without randomization, to CART BP pro-guided titration (CART BP arm, n = 67) or usual care (Control arm, n = 179). The primary endpoint was 24-hour mean systolic BP at 6 months, analyzed using analysis of covariance with 10 prespecified covariates. Sensitivity analyses addressed measured confounding, missing outcomes, and visit frequency. RESULTS: Among 162 participants with a valid 6-month ABPM recording, the adjusted between-group difference in 24-hour systolic BP was -7.35 mmHg (95% confidence interval [CI], -12.03 to -2.67 mmHg; P < 0.01), favoring the CART BP arm. Sensitivity analyses for measured confounding, missing outcomes, and visit frequency yielded similar estimates. In exploratory phenotype-specific comparisons, the controlled-BP phenotype was more prevalent in the CART BP arm (65.8% vs. 36.3%; P < 0.01), and masked hypertension was less prevalent (5.3% vs. 20.2%; P = 0.04). Adjusted office systolic BP differed by -3.14 mmHg (95% CI, -9.25 to 2.96; P = 0.31). Adverse events were more frequent per participant (20.0% vs. 8.1%; P = 0.02); exposure-adjusted incidence rate ratios were 1.78 (95% CI, 0.87 to 3.65) per participant-visit and 2.75 (95% CI, 1.35 to 5.61) per person-year. CONCLUSIONS: In this non-randomized study, a care pathway incorporating cuffless 24-hour ABPM was associated with lower 24-hour ambulatory systolic BP at 6 months compared with usual care, whereas office systolic BP did not differ significantly between arms. Given the non-randomized design, these findings are hypothesis-generating and require confirmation in a randomized trial. TRIAL REGISTRATION: ClinicalTrials.gov Identifier: NCT07454148.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Real-world performance of a ring-type cuffless blood pressure device compared with ambulatory blood pressure monitoring.Clinical hypertension · 2026
- Validation of ring-type cuffless blood pressure (BP) monitoring device for detecting awake-asleep BP variations compared to ambulatory BP monitoring device.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Nocturnal Blood Pressure Dipping in Adults: A Comparison Between Cuffless and Ambulatory Blood Pressure Monitoring.American journal of hypertension · 2026
- Continuous assessment and research on enhanced blood pressure control - CareBP study.Frontiers in cardiovascular medicine · 2026
- Evaluation of Alysis-001 cuffless blood pressure estimation algorithm against the European Society of Hypertension awake/asleep test criteria in hypertensive patients.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring.Journal of hypertension · 2026