Validation of ring-type cuffless blood pressure (BP) monitoring device for detecting awake-asleep BP variations compared to ambulatory BP monitoring device.
Seung-Mok Lee, Hyuktae Kwon, Belong Cho, Jin Ho Park, Sang-Hoon Na, Seung-Pyo Lee, Hae-Young Lee
Division of Cardiology, Department of Internal Medicine, Seoul National University Hospital, Seoul, Republic of Korea. · Department of Family Medicine, Seoul National University Hospital, Seoul National University College of Medicine, Seoul, Republic of Korea.
Key points
From the paper's conclusion- We evaluated the accuracy of a ring-type cuffless BP device (CART BP Pro) in detecting awake-asleep BP changes compared to conventional 24-h ambulatory BP monitoring (ABPM), following the validation protocol recommended by the European Society of Hypertension (ESH).
Abstract
We evaluated the accuracy of a ring-type cuffless BP device (CART BP Pro) in detecting awake-asleep BP changes compared to conventional 24-h ambulatory BP monitoring (ABPM), following the validation protocol recommended by the European Society of Hypertension (ESH). Among 84 recruited patients, 35 were selected to satisfy ESH recommendations: 30% (n = 11) with systolic BP (SBP) ≥ 130 mmHg and diastolic BP (DBP) ≥ 80 mmHg and 30% (n = 11) with SBP < 120 mmHg and DBP < 70 mmHg. The remaining 40% (n = 13) were included sequentially based on recruitment order. CART BP Pro was calibrated at least one day prior to simultaneous monitoring with ABPM. Participants simultaneously wore both devices on the non-dominant arm for 24 h. Mean SBP/DBP values and awake-asleep BP changes were compared. Awake-asleep SBP/DBP changes were 15.61 ± 8.74/11.35 ± 6.22 mmHg (ABPM) and 13.87 ± 9.04/11.44 ± 8.62 mmHg (CART BP Pro). Differences in mean awake-asleep BP changes were -1.74 ± 6.47 mmHg (SBP) and 0.08 ± 5.84 mmHg (DBP). Correlation analysis showed clinically significant correlation for both SBP and DBP changes (r = 0.736, p < 0.01). Bland-Altman plots showed acceptable agreement with a mean difference and 95% limits of agreement of -1.74 (-14.42, 10.94) mmHg for SBP and 0.08 (-11.37, 11.53) mmHg for DBP. For four-category dipping classification, CART BP Pro showed moderate agreement with ABPM (weighted κ = 0.58). In conclusion, CART BP Pro showed a meaningful level of agreement with ABPM in detecting awake-asleep BP changes. As this addressed only the awake/asleep test of the ESH protocol, the device cannot yet replace conventional 24-h ABPM pending completion of the remaining validation tests.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Real-world performance of a ring-type cuffless blood pressure device compared with ambulatory blood pressure monitoring.Clinical hypertension · 2026
- Evaluation of Alysis-001 cuffless blood pressure estimation algorithm against the European Society of Hypertension awake/asleep test criteria in hypertensive patients.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Nocturnal Blood Pressure Dipping in Adults: A Comparison Between Cuffless and Ambulatory Blood Pressure Monitoring.American journal of hypertension · 2026
- Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring.Journal of hypertension · 2026
- Continuous 24-hour cuffless blood pressure monitoring reveals high prevalence of non-dipping and blunted nocturnal heart rate reduction in rate-controlled atrial fibrillationEuropean Heart Journal · 2026
- Cuffless 24-hour ambulatory blood pressure monitoring-guided antihypertensive titration in real-world clinical practice: a prospective non-randomized study.Clinical hypertension · 2026