Continual cuffless blood pressure monitoring in children: agreement and feasibility compared with ambulatory blood pressure monitoring.
Michal Huml, Alexandra Kotková, Tereza Fremuthová, Karel Fiklík, Renata Pomahačová, Stanislav Kormunda, Josef Sýkora, Jiří Fremuth
Department of Pediatrics, University Hospital in Pilsen and Faculty of Medicine in Pilsen, Charles University, Pilsen, Czech Republic.
Key points
From the paper's conclusion- In this selected pediatric cohort, continual cuffless BP monitoring was feasible and showed group-level comparability with ABPM for averaged 24-h BP values. However, it should not be considered interchangeable with ABPM for individual clinical decisions or assessment of nocturnal dipping. Larger studies in more diverse pediatric populations are required before broader clinical implementation.
Abstract
BACKGROUND: Cuffless wearable technologies enable continual blood pressure (BP) monitoring and may improve tolerability compared with conventional cuff-based methods. Pediatric data comparing cuffless devices with ambulatory blood pressure monitoring (ABPM) remain limited. METHODS: In this prospective observational study, children with well controlled type 1 diabetes mellitus and normal office BP underwent simultaneous 24-h monitoring using a wrist-worn cuffless optical device and conventional ABPM. Between-method differences were assessed for averaged 24-h, daytime, and nighttime systolic and diastolic BP. Equivalence was evaluated using the two one-sided tests procedure with a prespecified exploratory ± 10 mmHg margin, whereas individual-level agreement was assessed using Bland-Altman analysis. RESULTS: Twenty-one participants were included (mean age 14.7 ± 2.6 years). Mean differences (Aktiia - ABPM) were -4.5 mmHg for 24-h systolic BP and -0.8 mmHg for 24-h diastolic BP. The 90% confidence intervals were fully contained within the ±10 mmHg margin for 24-h and nighttime systolic BP and for diastolic BP across all periods, but not for daytime systolic BP. Bland-Altman analysis showed modest mean bias but wide limits of agreement, particularly for systolic BP. Nocturnal systolic and diastolic BP dipping was attenuated with cuffless monitoring. CONCLUSIONS: In this selected pediatric cohort, continual cuffless BP monitoring was feasible and showed group-level comparability with ABPM for averaged 24-h BP values. However, it should not be considered interchangeable with ABPM for individual clinical decisions or assessment of nocturnal dipping. Larger studies in more diverse pediatric populations are required before broader clinical implementation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Nocturnal Blood Pressure Dipping in Adults: A Comparison Between Cuffless and Ambulatory Blood Pressure Monitoring.American journal of hypertension · 2026
- Real-world performance of a ring-type cuffless blood pressure device compared with ambulatory blood pressure monitoring.Clinical hypertension · 2026
- Validation of ring-type cuffless blood pressure (BP) monitoring device for detecting awake-asleep BP variations compared to ambulatory BP monitoring device.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Long-Term Stability and Blood Pressure Change Tracking with a Cuffless Wearable Device in Hypertensive Patients.Blood pressure · 2026
- Accuracy of cuffless blood pressure devices compared with invasive arterial line monitoring in high-acuity settings: a random-effects meta-analysisEuropean Heart Journal · 2026
- Evaluation of Alysis-001 cuffless blood pressure estimation algorithm against the European Society of Hypertension awake/asleep test criteria in hypertensive patients.Hypertension research : official journal of the Japanese Society of Hypertension · 2026